16
How to replace the medication chamber
Attention
Attention
The medication chamber is a maintenance part and does not carry any warranty.
Under normal conditions, the lifetime of the medication chamber is approximately
6 months (three times usage per day or 30 minutes per day). However, the
nebulization performance may start deteriorating in less than 6 months depending
on the way you use it or the use of certain types of medication. If the nebulizer can
not nebulize or the nebulization rate decreases significantly after clean, you must
replace the medication chamber with a new one. (If you want to purchase a
medication chamber, please contact the retailer from which you purchased the
product or any nearby retailers.)
1. Remove
the
medication
chamber from the nebulizer:
Press the PUSH button on the rear
side of the main unit, and push the
medication chamber toward the
front side of the main unit.
2. Re-attach
the
medication
chamber to the nebulizer
Attach the medication chamber
correctly as shown in the figure.
• To avoid damaging the nebulizer,
please ensure the PUSH button is
pressed down before pushing the
medication chamber forward.
• To avoid rupturing the mesh,
please do not poke it with your
finger or other foreign object.
• Please ensure that the medication
chamber is attached correctly;
otherwise, it may result in a bad
connection
and
cause
the
nebulizer to not function properly.
• Please keep the electrodes of the
main unit and medication chamber
clean; otherwise, the nebulizer
may not function properly.
• Please clean the medication
chamber before using.
Press
Push forward