Technical Data | 10
Instruction Manual VACTEST DCC 400_EN_en
19 / 24
10 Technical Data
VacTest DCC 400
Measurement principle
Inverted magnetron
Materials exposed to vacuum
Stainless steel 1.4307, nickel, tungsten, molybdenum, glass,
ceramic
Measuring range
mbar
2 x 10
-3
… 5 x 10
-9
torr
1.5 x 10
-3
… 3.75 x 10
-9
Overpressure limit
bar abs.
10
Measurement uncertainty
% of reading
2 x 10
-3
… 1 x 10
-8
mbar:
±
25 %
Repeatability of measurement
% of reading
2 x10
-3
… 1 x 10
-8
mbar:
±
5 %
Leakage rate
mbar l/s
< 5 x 10
-10
Reaction time
ms
< 50
Serial Interface
RS485
Electrical Connection
D-Sub, 15-pin, male
Supply voltage
V
20 … 30
Cathode voltage
V
2.5
Max. power consumption
W
normal operation ► 3
W
relays/display ► 0.8
Output signal
0 … 10 V, RS485
Output function
V
log(p
mbar
) + 10.5
Under-range voltage
V
0.9
Error
V
0.4
Setpoint relay
2 dry contacts
Relay contact rating
2A, 50 VAC / 2 A, 30 VDC, max. 60 VA
Operating Temperature
°C
+5 ... +60
Storage Temperature
°C
-40 ... +65
Max. bake-out temperature at the
flange
°C
160
Relative Humidity:
max. 80% up to 30°C, max. 50% at 40°C, not condensing
Ambient pressure:
hPa (mbar)
860 … 1060
Protection Class
IP 40 (IP 54 with appropriate D-Sub connector)
Weight
g
555