Seco SG 0008 A
0870162612 /
190709
page 11
The vacuum pump is blocked
Solid foreign matter has entered the vacuum
pump
Repair the vacuum pump (Busch service)
Clean the suction line
If necessary additionally provide a filter
The vacuum pump was run in the wrong direc-
tion
Repair the vacuum pump (Busch service)
When connecting the vacuum pump make sure
the vacuum pump will run in the correct direc-
tion (
➔
page 6 Installation)
The drive motor is running, but the vacuum
pump stands still
The coupling between the drive motor and the
vacuum pump is defective
Replace the coupling element (Busch service)
The vacuum pump runs very hot (thermal
protection switch triggers)
Insufficient air ventilation
Make sure that the cooling of the vacuum
pump is not impeded by dust/dirt
Install the vacuum pump in a narrow space
only if sufficient ventilation is ensured
Ambient temperature too high
Observe the permitted ambient temperatures
Temperature of the inlet gas too high
Observe the permitted temperatures for the in-
let gas