8
Troubleshooting
Symptom
Potential Cause
Solution
Battery pack will not connect
properly to handpiece.
There is debris in the battery receptacle.
See cleaning recommendations.
Battery pack contacts are damaged.
Replace battery pack.
Battery pack latch is damaged.
Replace battery pack.
Battery receptacle is damaged.
Send handpiece to Brasseler U.S.A. Medical, LLC, for service.
Battery pack malfunctions.
Over sterilization of battery pack.
Replace battery pack.
End of life cycle for battery pack.
Replace battery pack.
Subjected to a dry cycle.
Replace battery pack.
4-Bay Power Unit fuses are open/blown.
Replace with appropriate rated fuses. See 4-Bay
Power Unit instructions for use.
Battery pack cannot be
removed from handpiece.
Battery pack latch is damaged.
Send handpiece and battery pack to Brasseler U.S.A.
Medical, LLC, for service.
Battery receptacle is damaged.
Send handpiece and battery pack to Brasseler U.S.A.
Medical, LLC, for service.
Battery pack is fully charged but
won’t run.
Battery pack malfunctioned.
Replace battery pack.
Handpiece malfunctioned.
Send handpiece to Brasseler U.S.A. Medical, LLC, for service.