SURGICAL POWER & ACCESSORIES

Summary of Contents for BSP MAX II PM-X00-710

Page 1: ...PM X00 710 PM X00 715 IR 12 13 SURGICAL POWER ACCESSORIES...

Page 2: ......

Page 3: ...planation of Symbols Page 6 Features Page 6 Battery Pack Installation Removal Page 7 Troubleshooting Page 8 Cleaning Recommendations Page 9 Sterilization Recommendations Page 10 Specifications Page 12...

Page 4: ...care cleaning sterilization and long term maintenance of BSPMAX II Large Bone Power System Introduction The BSPMAX II 9 6V Battery Pack PM X00 710 and 12V Battery Pack PM X00 715 are intended for use...

Page 5: ...es Consult WHO and local regulations for further information Battery packs contain nickel cadmium which must be recycled or disposed of properly See product disposal instructions DO NOT store battery...

Page 6: ...release the battery pack from the handpiece Battery Pack Latch Battery Pack Case Authorized representative in the European community Batch code Catalog number Caution Conforms with the essential requ...

Page 7: ...battery pack is secure Rotate trigger to the RUN mode Test the handpiece and battery pack by depressing the trigger Rotate trigger to the SAFE mode when not in operation Battery Pack Removal Rotate t...

Page 8: ...ttery pack End of life cycle for battery pack Replace battery pack Subjected to a dry cycle Replace battery pack 4 Bay Power Unit fuses are open blown Replace with appropriate rated fuses See4 Bay Pow...

Page 9: ...battery pack contacts with abrasives Dry battery packs before charging to avoid damage to the battery packs and charger Cleaning Procedures 1 Remove battery pack from the handpiece 2 Inspect the batte...

Page 10: ...sterilize battery packs in cold sterilization like CIDEX DO NOT Peel Pack battery packs for sterilization Sterilization in a sealed pouch traps moisture which can cause damage DO NOT sterilize batter...

Page 11: ...MAX II battery packs The sterilizer manufacturer s written instruction for cycle parameters load configuration and AAMI guidelines for steam sterilization should be followed Sterilization Recommendati...

Page 12: ...ully Charged 3 10 Minutes 3 15 Minutes Working Temperature 32 F 122 F 0 C 50 C 32 F 122 F 0 C 50 C Type Nickel Cadmium Ni Cd Nickel Cadmium Ni Cd Specifications Operating Ambient temperature 50 F to 7...

Page 13: ...13 Return Goods Policy Contact your distributor regarding returned goods policy...

Page 14: ...4 IN KEYLESS CHUCK PM X08 935 3 MM KEYLESS CHUCK PM X08 940 ZHS ZIMMER HUDSON STRYKER UNIVERSAL ATTACHMENTS REAMERS PM X08 911 1 4 IN JACOBS HI TORQUE W KEY EXTENDED LENGTH PM X08 950 HUDSON EXTENDED...

Page 15: ......

Page 16: ...stered trademark of Hudson Industries Inc Jacobs is a registered trademark of Jacobs Chuck Manufacturing Company Stryker is a registered trademark of Stryker Corporation Zimmer is a registered tradema...