4.4
Mattress base sizes
This instruction manual may describe functions or features that your model of bed does
not have.
The switch mode power supply consists of: Voltage transformer and low-voltage con-
nection cable and can be used as instruments to separate the bed from the power sup-
ply system.
The voltage transformer generates a protective low voltage that is safe for both the res-
ident and care staff.
The switch mode power supply provides all drives (motors) with protective low voltage
using a connection cable. The socket available is protected against the ingress of wa-
ter.
•
A central control unit. All drive motors and the handset (Bluetooth handset) are cou-
pled/connected to the central control unit;
•
A Bluetooth handset with a sturdy hook (connected to the control unit via a Bluetooth
interface), for Dali Standard and Dali wash.
The user can lock the adjustment options on the handset if the poor clinical condition of
the resident necessitates this.
•
A cabled handset with a sturdy hook, for Dali Standard and Dali low entry
The user can lock the adjustment options on the handset if the poor clinical condition of
the resident necessitates this.
•
Two drive motors for horizontal height adjustment.
•
A drive motor for the thigh rest.
•
A drive motor for the backrest.
And for Dali econ and low-econ, low-entry-econ:
•
A central control unit. All drive motors and the cabled handset are connected to the
central control unit;
•
A mains cable: a spiral cable with EPR sheathing and country-specific plug. These
plugs can be used as instruments to separate the bed from the power supply system.
The mains cable is interchangeable by maintenance personnel.
•
A cabled handset with sturdy hook connected to the control unit via plug connection.
The user can lock the adjustment options on the handset if the poor clinical condition of
the resident necessitates this.
•
Two drive motors for horizontal height adjustment.
•
A drive motor for the thigh rest.
•
A drive motor for the backrest.
The Dali care bed can be ordered in the following sizes.
Product Description
8
Part A: General Information
Summary of Contents for Dali
Page 2: ...Part A General Information ...
Page 4: ......
Page 6: ...II ...
Page 30: ...Part B Operator and Technical Per sonnel ...
Page 32: ......
Page 36: ...IV ...
Page 88: ...Part C Care staff and residents ...
Page 90: ......
Page 129: ...Cleaning and Disinfection Part C Care staff and residents 37 ...
Page 130: ...Cleaning and Disinfection 38 Part C Care staff and residents ...
Page 131: ...Cleaning and Disinfection Part C Care staff and residents 39 ...