6.9
Electrical specifications of the
8619 multiCELL WM AC
Electrical supply 110...240 V AC
• Frequency
• Max. current
• Integrated protection
• 50...60 Hz
• 500 mA
• delayed 3.15 A fuse, 250 V AC, (breaking capacity = 1500 A at
250 V AC, 10 kA at 125 V AC), certified EIC 60127, UL listed
and recognized
Power distribution
(POWER OUT module)
• Protected against polarity reversal
• 24 V DC, filtered and regulated, SELV circuit, at a non-hazardous
energy level,
•
1,3 A max.: observe the maximum permissible load as a
function of the ambient temperature. See the derating curves
Fig. 10.
8619 multiCELL WM AC,
without extension module
8619 multiCELL WM AC,
with extension module
Maximum current
of the load
Ambient temperature
[A]
[°C]
-20
0
-10
+10 +20 +30 +40 +50 +60 +70
0
0.6
1.1
0.1
+80
0.8
1.3
Fig. 10 : Derating curves of the maximum permissible current, depending on the ambient temperature
All digital inputs ("DI")
• Switching threshold V
on
: 5...36 V DC
• Switching threshold V
off
: < 2 V DC
• Input impedance: 3 k
W
• Galvanically isolated
• Protected against polarity reversal and voltage spikes
• Frequency: 0.5...2500 Hz
32
Technicaldata
Type 8619
English
Summary of Contents for 8619 multiCELL WM AC
Page 6: ...6 English ...
Page 12: ...12 Type 8619 English ...
Page 24: ...24 Type 8619 English ...
Page 40: ...40 Type 8619 English ...
Page 70: ...70 Installation and wiring Type 8619 English ...
Page 180: ...180 Type 8619 English ...
Page 192: ...192 Type 8619 English ...
Page 193: ......
Page 194: ...www burkert com ...