217
SAW sensor – Parameter
Type 8098 FLOWave L
15.12.3
Activating the damping of the differentiation factor values and
choosing a predefined damping level
The damping makes it possible to damp the fluctuations of the measured values of the DF:
• on the display,
• on the totalizers,
• on the outputs. The damping set for an analog output comes in addition to the damping of the DF.
The damping is not applied to the new measured value, if the 2 following conditions are met:
• a
Low
,
Medium
or
High
damping level is active,
•
and the variation between 2 values that are measured one after the other is higher than 30 % (for example
when changing the liquid in the pipe).
By default, the DF values are damped with the level
None
.
The
Low
damping level or no damping at all (
None
) are suited for applications/processes that need fast
response times.
The
Medium
damping level or the
High
damping level are suited if the DF values change slowly.
→
As an alternative to the 3 predefined damping levels
Low
,
Medium
or
High
, you can set your own
damping parameters. See chapter 15.12.4
.
None
damping: all
the fluctuations of the
measured values are
transmitted
Low
damping
Medium
damping
High
damping
Chosen damping level is applied
No damping if 2 consecutive measured values vary for ±30 %
Fig. 61:
Operation of the available damping levels
Table 30:
Response times (10 %...90 %) of the damping levels for the DF measurements
Damping level
Response time
None
0 s
Low
1 s
Medium
10 s
High
30 s
Special
User-defined
Response time
: see chapter 15.12.4
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...