Pressure Transmitter 8326
Operation of Transmitters with Displays
Subject to technical changes without notice.
GB-29
Example:
Level 100 %:
4000 mm
Density:
1 g/cm3
Density correction:
0.9 g/cm3
Span end point:
The span (end-point) must be re-calibrated (with or without pressure) to 4000 mm in
order to prevent a 4000 mm level tank from being overfilled.
6.5.5
Main Menu: Language
A change or correction in the density causes a change in the span end-
point’s unit of measure (mm, m, inch, feet). The span end-point must be
re-calibrated when changing the medium to be measured (due to a
change in density).
4000 mm 1 g/cm
3
⋅
0.9 g/cm
3
--------------------------------------------------
4444
mm
=
Important