7.10. product with a pg13.5 connection
onto a hygienic holder with clamp
connection
• To avoid any mechanical damage to O-rings during
assembly, lightly grease them.
• After assembly, clean the product of any remaining drops of grease.
1
2
(not delivered)
→
Make sure the product and the holder are not damaged.
→
Make sure all O-rings are in place in their appropriate grooves,
and are not damaged.
→
Make sure the holder 2 is fixed onto the clamp of the pipe or
onto the clamp of the vessel.
→
Screw the product 1 in the holder 2.
Fig. 15: Installation of a 8221 with a PG13.5 connection onto a
holder with clamp connection onto a pipe or on a vessel
8.
Wiring
8.1.
product with 4 electrodes and a
cable, process connection 1½''
clamp or g1¼''
signal description
cable colour
bürkert controller
type 8619
Pt1000 (low end)
grey
7 SE
Pt1000
white
8 TS
Pt1000
blue
9 TS
Current electrode (high end)
pink
1 C+
Potential electrode (high
end)
green
2 P+
Potential electrode (low end)
brown
3 P–
Current electrode (low end)
yellow
4 C–
Not connected on product
shield
6 FE
Not connected
red
-
26
Wiring
Type 8221
English