background image

Page 17

1062DOC  Rev C

Effective date: 6-Mar-14

RAMVAC Bulldog

®

 QT

User Guide

Problem

Possible Cause

Corrective Action*

Low or No Vacuum

(motor running okay)

Tank Drain Valve Blocked Open

Clogged Vacuum Line

Filtrols Check Valve Not Sealing

Vacuum Leaks

Loose or Broken Drive Belt

Stuck Vanes

Clean Tank Drain Valve

Locate and remove clog

Repair or replace Check Valve

Locate and fix leaks

Tighten or replace Belt

Clean and lubricate Vanes

Motor Does Not Run

“Tripped” Breaker or Fuse

“Tripped” Motor Overload

Motor Failure

Failed Control Component

Filtrols Moisture Fault

Reset Breaker / Replace Fuse

Reset Motor Overload

Replace Motor

Bypass then replace failed component

See “Filtrols Moisture Fault”

Drip Rate Slow

Low Ambient Temperature

Dirty Oil Filter

Improper Oil

Raise ambient temperature

Clean Filter

Change to recommended oil

Drip Rate Fast

High Ambient Temperature

High Vacuum Pressure

Improper Oil

Lower ambient temperature

Lower vacuum pressure

Change to recommended oil

Oil Comes Out Exhaust

Incorrect Exhaust Installation

Oil Reservoir Overfull

Bypassed Moisture Fault

Correct Installation

Lower oil level

See “Filtrols Moisture Fault”

“Filtrols Moisture “ Fault

Separating Tank overfilled

Separating Tank Drain Valve Stuck

Foaming Line Cleaner used

Cold Sterilization Solution aspirated

Drain Separating Tank

Clean Separating Tank Drain Valve

Use “Slugbuster” Line Cleaner

Aspirate only non-foaming substances

Maintenance Required

Preventative Maintenance Due

Perform Preventative Maintenance

Troubleshooting

*Abbreviated information.  For details contact your authorized dealer or RAMVAC, or refer to the “Support” section of ramvac.com

Summary of Contents for RAMVAC Bulldog QT

Page 1: ...OPERATION MAINTENANCE AND USER MANUAL DENTAL VACUUM SYSTEM...

Page 2: ...s and Cautions Thank you for selecting RAMVAC to serve your dental facility RAMVAC provides industry leading technology for convienient and efficient utility room equipment Invest a few minutes of you...

Page 3: ...l concentrations typically encountered in dental analgesia The additional air drawn into a properly installed and operated Vacuum Unit will dilute these agents Never use your RAMVAC to remove pure nit...

Page 4: ...N Guidance and manufacturer s declaration electromagnetic immunity The Bulldog is intended for use in the electromagnetic environment specified below The customer or the user of the Bulldog should ass...

Page 5: ...tromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz Does Not Comply Does Not Comply Portable and mobile RF communicat...

Page 6: ...9 Maintenance Maintenance Overview Page 10 Preventive Maintenance Schedule Page 10 Oil Fill Level Page 10 Cleaning the Vacuum System Page 11 Air Filters Page 11 Lubrication System Components Page 12 O...

Page 7: ...ays or be otherwise suitable for motor circuit 3 Disconnect must be fusible and sized per UL 98 UL 489 or UL 508 Must be supplied and installed by a licensed electrician 4 Motors are NEMA rated Electr...

Page 8: ...Page 8 1062DOC Rev C Effective date 6 Mar 14 User Guide RAMVAC Bulldog QT MOTOR L1 MOTOR L2 LINE IN L1 LINE IN L2 GROUND Electrical Connection S2 Logic Board...

Page 9: ...chematic Gases Continue To Pump Facility Vacuum Piping Separating Tank Liquids and Solids Drain When Vacuum is off Bulldog Vacuum Unit Power To Motor Exhaust To Outside Weather Cap Required Remote Con...

Page 10: ...ally Remote switch indicator will flash slowly S2 and or OWL Touch will indicate maintenance is due on the display Perform Preventative Maintenance Hold reset button until it resets Moisture in Filtro...

Page 11: ...1 000 hours Check Air Filters See Air Filters Check Drip Rate See Check Oil Drip Rate Every 2 000 hours Change Oil and check oil filter See Oil Change Check V Belt See Check V Belt The S2 will display...

Page 12: ...unces of an appropriate vacuum line cleaner such as SlugBuster through the lines For overhead plumbing be sure to allow air to follow liquids before closing vacuum valves Clean treatment room solids s...

Page 13: ...is its Lubrication System The Lubrication System supplies oil drop by drop through Drippers to the pump Used oil is discharged back into the Oil Reservoir as a mist separated from the exhaust filtered...

Page 14: ...ip rate See Check Oil Drip Rate Dispose of used oil at a gas station or lubricant recycling station 4 Reset Change Oil From Change Oil screen press and hold the reset button for 5 seconds Check Oil Dr...

Page 15: ...trol Terminals Toggle Switch or other non illuminated switch option Air Switch Air Light Common F 24 VDC G 24 VDC H DC Common Illuminated Remote Switch Note Maximum wire length for low voltage 18 gaug...

Page 16: ...sture Allow the Separating Tank to drain Lift off the Vacuum Controller remove Main Air Filter Clean out any remaining moisture Remove Moisture Sensor by twisting and pulling down and dry contacts See...

Page 17: ...nt Temperature Dirty Oil Filter Improper Oil Raise ambient temperature Clean Filter Change to recommended oil Drip Rate Fast High Ambient Temperature High Vacuum Pressure Improper Oil Lower ambient te...

Page 18: ...RAMVAC Dental Vacuum System 2 Years from date of installation RAMVAC Vacuum Pumps only 10 Years from date of installation RAMVAC OWL 2 Years from date of installation CustomAir by RAMVAC 6 Years 4200...

Page 19: ...Page 19 1062DOC Rev C Effective date 6 Mar 14 RAMVAC Bulldog QT User Guide...

Page 20: ...www dentalez com www ramvac com 1062DOC Rev D Effective date 6 May 17 2012 RAMVAC Dental Products All rights reserved No part of this publication may be copied or distributed transmitted or transcrib...

Reviews: