GAS 222.20 DH
2.2 General Hazard Warnings
The maximum surface temperatures of the probes also vary based on operating conditions
(steam temperature, sample gas in-
let temperature, ambient temperature, fluid flow rate). When used in
explosive areas, also particularly note
the related hazard
warnings.
The equipment must be installed by a professional familiar with the safety requirements and risks.
Be sure to observe the safety regulations and generally applicable rules of technology relevant for the installation site. Prevent
malfunctions and avoid personal injuries and property damage.
The operator of the system must ensure:
– Safety notices and operating instructions are available and observed,
– The respective national accident prevention regulations are observed,
– The permissible data and operational conditions are maintained,
– Safety guards are used and mandatory maintenance is performed,
– Legal regulations are observed during disposal,
– compliance with national installation regulations.
Maintenance, Repair
Please note during maintenance and repairs:
– Repairs to the unit must be performed by Bühler authorised personnel.
– Only perform conversion-, maintenance or installation work described in these operating and installation instructions.
– Always use genuine spare parts.
– Do not install damaged or defective spare part. If necessary, visually inspect prior to installation to determine any obvious
damage to the spare parts.
Always observe the applicable safety and operating regulations in the respective country of use when performing any type of
maintenance.
NOTICE
When used in explosive areas
The basic version of the probe is not suitable for use in
Zone 1 or 21
explosive areas. The
accessories selected (e.g. add-on parts such as solenoid valves, heated blowback vessel...)
may severely limit the approved applications.
When using
category 3G or 3D accessories
, the are of application for the probe is limited
to
Zone 2 or Zone 22
. Therefore be sure to note the type plates for all add-on parts. Partic-
ularly pay attention to the ignition protection markings and all hazard warnings in this
manual.
DANGER
Toxic, corrosive gas/condensate
Sample gas/condensate may be hazardous to health.
a) If necessary, ensure a safe gas/condensate discharge.
b) Always disconnect the gas supply when performing maintenance or repairs.
c) Protect yourself from toxic/corrosive gasses/condensate when performing mainten-
ance. Wear appropriate protective equipment.
DANGER
Explosion hazard
Life and explosion risk may result from gas leakage due to improper use.
a) Use the devices only as described in this manual.
b) Regard the process conditions.
c) Check tubes and hoses for leakage.
4
Bühler Technologies GmbH
BE460045 ◦ 10/2019
Summary of Contents for GAS 222.20 DH
Page 20: ...GAS 222 20 DH 9 3 Dimensions 18 Bühler Technologies GmbH BE460045 10 2019 ...
Page 23: ......
Page 24: ......
Page 32: ...DE 461099 05 2021 Page 8 9 ...