EGK 4S
4.2.1 Connecting the heat exchanger
The gas inputs are marked in red.
On glass heat exchangers the correct position of the seal is important when connecting the gas lines (see image). The seal con-
sists of a silicone ring with a PTFE sleeve. The PTFE side must face the glass thread.
Glass
PTFE
4.3 Electrical connections
The operator must install an external separator for the device which is clearly assigned to this device.
This separator
– must be located near the device,
– must be easy for the operator to reach,
– must comply with IEC 60947-1 and IEC 60947-3,
– must separate all live conductors and the status output, and
– must not be attached to the power feed.
WARNING
Hazardous electrical voltage
The device must be installed by trained staff only.
CAUTION
Wrong mains voltage
Wrong mains voltage may damage the device.
Regard the correct mains voltage as given on the type plate.
WARNING
High voltage
Damage to the device in case of insulation testing
Do not proceed insulation tests with high voltage
to the device as a whole!
Insulation test
The device is equipped with extensive EMC protection. If insulation tests are carried out the electronic filter devices will be dam-
aged. All necessary tests have been carried out for all concerned groups of components at the factory (test voltage 1 kV or 1.5 kV
respectively, depending on the device).
If you wish to carry out the insulation test by yourself, please test only separate groups of components.
Disconnect the compressor, the fan, the heating or the peristaltic pumps, respectively, and then carry out the insulation tests.
8
Bühler Technologies GmbH
BE450004 ◦ 10/2018