BA 4510
NOTICE
When installing
connections
for steel lines, the information in the appendix provided by
the manufacturer must be observed.
Silicone tubes
can cause inaccurate measurements due to their oxygen permeability.
The manufacturer therefore advises against using this type of tubing.
CAUTION
Water entering the hot measuring cell
Water entering the hot measuring cell can damage it beyond repair and must therefore
be avoided.
If the sample gas contains so much water vapour there is a risk of water condensation,
inside cold tubing, a water separator must be installed before the sample gas enters the
device.
NOTICE
The sample gas can also flow with the device switched off.