5.2 Operating the sample gas pump
CAUTION
Risk of injury by moving parts
The device may be damaged if it falls down of by impacts. Pay attention to any accessible
moving parts.
Operation without cover or with damaged cover is not allowed!
The sample gas pump is suitable for delivering gases only. It is not suitable for liquids.
The sample gas pump should be operated without pressure. A system pressure above 0.3 is not allowed. The gas outlet must not
be closed. The flow rate must be at least 50 l/h (150 l/h with system pressure 0.3 bar). If the flow rate continuously is throttled be-
low 150 l/h during operation, the flow rate must be controlled with a bypass valve.
NOTICE
Extreme throttling reduces the life time of the bellow.
If the pump is equipped with a bypass valve, the flow rate can be adjusted. Do not apply force when turning the valve, because
the valve may be damaged otherwise! The turning range of the valve is about 5 turns.
14
Bühler Technologies GmbH
BE420020 ◦ 03/2020