Nivotemp NT M-XP
5.7.3 Switching output x: Lower switching limit (switch-back point)
The lower switching limit for switching output Out x can be defined with the following submenu:
Measured variable
____
out_
Measured variable main menu
Out X
____
rPI FLI F I.Lo
Switch-back point X
____
Selection menu
[ o.Lo - o.Hi ]
Settings menu OUTX
Setting range:
[
o.Lo
]…[
o.Hi
]
Switch-back point for OUT x
Note:
– The switch-back point must be set to within the range limits.
– If switching output OUT x was assigned the function
Window
,
will appear. The
setting corresponds with the lower window limit.
– If switching output OUT x was assigned the function
Frequency output
,
will ap-
pear. The setting corresponds to the frequency 1 Hz.
5.7.4 Switching output x: Switch-on delay
The menu
Advanced Functions
EF
x
is used to configure additional settings for switching output x. The submenu is at the second
submenu level.
The switching and switch-back delay prevents the alarm being triggered too frequently in unstable conditions. The switching
delay can be configured with the following menu:
Measured variable
____
out_
Measured variable main menu
Out X
EF_
EF X
Advanced Functions
dS_
Delay SP X
Delay OUTX On
___
Selection menu
0 - 100
Settings menu OUTX
Setting range:
0…100 seconds
Time span in seconds during which the signal must be continuously present for the switching out-
put to respond.
Note:
– If switching output OUT x was assigned the function
Window
, the setting corresponds to the
switch-on delay which detects valid reaching of the measuring window.
– If switching output OUT x was assigned the function
Frequency output
, this value will have no
affect.
5.7.5 Switching output x: Switch-back delay
The switch-back delay can be configured with the following menu:
Measured variable
____
out_
Measured variable main menu
Out X
EF_
EF X
Advanced Functions
dr_
Delay RP X
Delay OUTX Off
___
Selection menu
0 - 100
Settings menu OUTX
Setting range:
0…100 seconds
Switch-back signal delay for OUT x.
Time span in seconds during which the signal must be continuously present for the switching out-
put to respond.
Note:
– If switching output OUT x was assigned the function
Window
, the setting corresponds to the
switch-on delay which detects valid closing of the measuring window.
– If switching output OUT x was assigned the function
Frequency output
, this value will have no
affect.
20
Bühler Technologies GmbH
BE100023 ◦ 06/2021