EGK 1/2
Insulation test
The device is equipped with extensive EMC protection. If insu-
lation tests are carried out the electronic filter devices will be
damaged. All necessary tests have been carried out for all con-
cerned groups of components at the factory (test voltage 1 kV
or 1.5 kV respectively, depending on the device).
If you wish to carry out the insulation test by yourself, please
test only separate groups of components.
Disconnect the compressor, the fan, the heating or the peri-
staltic pumps, respectively, and then carry out the insulation
tests.
Connection via plug
The device is equipped with connectors according to EN
175301-803 for mains and status output. If the cables are
mounted properly, they cannot be interchanged. Please make
sure that the connectors are remounted correctly after con-
necting the cable. The following figures show the pin assign-
ment with respect to the numbers printed on the connector.
power supply
status output
pin assignment
function ok
alarm
Fig. 1:
A05-100002 Cooler electric supply
The mains supply must be protected with 10 A. The clamping
area has a diameter of 8-10 mm.
5 Operation and control
NOTICE
The device must not be operated beyond its specifications.
After switching on the cooler the block temperature will be
displayed. The display will flash until the block temperature
has reached the preset target value (± adjustable alarm
range). The status contact is in the Alarm position.
Once the target temperature range has been reached, the
temperature will continuously be displayed and the status
contact switches over.
If the display flashes during operation or an error message ap-
pears, please refer to bullet “Troubleshooting”.
Please refer to the data sheet for performance data and max-
imum ratings.
5.1 Use of menu functions
Overview of the operational principal:
Use this short description if you have experience with the
device.
Operation is carried out by only the keys with the following
functions:
Key
Function
• Switch from measurement display to main
menu
• Selection of the display menu item
• Accepting the changed value or selection
• Switch to the upper menu item
• Increase of the value of switching the selection
• Temporary display of the alternative measure-
ment display (if option is installed)
• Switch to lower menu item
• Decrease of the value of switching the selection
• Temporary display of the alternative measure-
ment display (if option is installed)
9
Bühler Technologies GmbH
BX450001 ◦ 10/2018