5. Make as many cross cuts through the dressing stick as needed to dress the blade.
Three to five crosscuts usually work well.
Lubricant Tank
Dressing Chuck with a Dressing Stick
Dressing Stick Feed Control Knob
Figure 3
Dressing Chuck inside the Lubricant Tank
WARNING! Personal Injury.
Feeding the dressing stick or any specimen
sample into the blade by hand can result in severe personal injury and equipment
damage.
Lubrication
The IsoMet
®
1000 Precision Saw uses the drag principle of lubrication. The lubricant is
carried to the sample on the edge of the rotating blade.
IsoCut
®
Plus Fluid reduces cutting time and produces superior cuts. IsoCut
®
Plus Fluid
is best for most metals and many non-metals. The use of IsoCut
®
Plus Fluid promotes
effective lubrication enables diamond particles to cut cleanly and minimizes loading of the
blade.
•
Fill the IsoMet
®
1000 lubricant tank with IsoCut
®
Plus Fluid to a level that will
immerse the blade approximately ¼-inch.
•
Discard the lubricant when becomes contaminated with sludge.
a. Remove the sample tray, wafering blade, and flanges from the arbor shaft.
b. From the front of the machine, pull the lubricant tank straight out.
c. Discard the lubricant in a safe and approved manner.
d. Clean the lubricant tank.
e. Reinstall the lubricant tank.
f. Refill the lubricant tank with fresh, clean lubricant.
g. Reinstall the wafering blade, flanges, and sample tray.
11
MA112180_33