Safety Warnings 

 

IMPORTANT:

 We strongly advise that 

individuals with pacemakers, cochlear 

implants, or trans-dermal electronic implants - 
e.g., glucose monitors do not use our high-
voltage test equipment under any 
circumstances. 

IMPORTANT:

 This product must not be 

used in combustible atmospheres or 

ATEX Hazardous Areas, such as fuel stations or 
in proximity to any combustible gas (and/ or 
oxygen concentrations exceeding normal 
atmosphere). 

All high-voltage testing equipment 
must be operated by suitably trained, 

responsible, authorised personnel only. 

IMPORTANT:

 This product can cause 

radio interference and caution must 

be taken when operating in sensitive 
environments such as hospital roofs. 

IMPORTANT:

 This instrument must be 

connected to an appropriate earth 

(ground) before a high voltage is applied. 

Observe all safety precautions; 
operators will be exposed to high 

voltages when using this instrument. Avoid 
contact with live components. 

This instrument creates sparks in use, 
which cause the formation of ozone 

(O

3

). Ozone is a known irritant and therefore - 

particularly when using the instrument in 
confined spaces - the user must ensure 
adequate ventilation so that established 
workplace exposure limits are not exceeded. 

Safety Warning:

 DO NOT USE this 

equipment in the event of it being 

immersed in water. It must be immediately 
returned to Buckleys for service, calibration 
and/or repair. 

A high voltage will always take the easiest route 
to earth. The risk of electrical shock is inherent 
when operating high-voltage equipment and 
extreme caution must be taken at all times. 

The output voltage of this equipment is derived 
from a high-impedance source and is current 
limited. However, when the equipment is in use, 
should the operator accidentally contact the 
test electrode they will receive an electric 
shock. If this situation is unavoidable, we 
recommend that suitable rubber or plastic 
insulating gloves be worn. Furthermore, the 
operator should be in good health and - in 
particular - not suffer from any cardiac 
condition. 

During testing, a static charge may accumulate 
on the surface under test. If the operator is 
walking on the surface, they may also become 
‘charged- up’. When discharged to earth, this 
charge could result in a static electric shock, 
such as may be experienced in the home under 
certain weather conditions. 

Whilst such a shock does not normally present 
any danger to the user, it can be unpleasant. In 
most cases, the operator can avoid this by 
wearing the blue electrostatic discharge 
wristband supplied and ensuring it is connected 
by the blue cable to the grounding terminal 
(4mm socket) on the test probe handle, or the 
earthing terminal on the instrument’s enclosure. 
In all cases it is essential to ensure that the 
instrument is correctly earthed

 

 

Summary of Contents for DRY ROOF PRO2

Page 1: ...Electronic Flat Roof Leak Detection and Location Kit Quick Start Guide Please read this guide before using theequipment 5830 0047 11 21 English...

Page 2: ...nings 3 EC Declaration of Conformity 4 UKCA Declaration of Conformity 4 Specifications 4 Maintenance 5 Register Your Product 5 Useful Advice 5 Kit Contents 6 Risk Assessment 7 Factory Settings 7 Setup...

Page 3: ...eing immersed in water It must be immediately returned to Buckleys for service calibration and or repair A high voltage will always take the easiest route to earth The risk of electrical shock is inhe...

Page 4: ...nge without notice or obligation Output voltage range 0 9kV 40kV Max output current 1mA continuous Alarm sensitivity range 10 A to 450 A factory set to 200 A Recommended coating thickness range 13 m 2...

Page 5: ...mmediately To register your product simply visit www buckleysinternational com registration complete the online form and click on SUBMIT Useful Advice Instrument sensitivity Several variables includin...

Page 6: ...5m earth cable on reel H D cell rechargeable batteries x4 I Battery charger with mains power adaptors and in car charger J Mid section extension rod x2 K End section extension rod L 450mm angled phosp...

Page 7: ...materials been removed from the area to be tested EMC RFI Buckleys high voltage instruments will produce sparks which may create interference in nearby electronic apparatus particularly communication...

Page 8: ...Quick Start Guide 8 Setup...

Page 9: ...croll to Alarm Suppression and press b Press d or e until the display shows the desired number and press b Press c to return to the main display screen Refer to page 5 for more information on sensitiv...

Page 10: ...Quick Start Guide 10 Basic Testing Press and hold the ON button 1m sec...

Page 11: ...11 Basic Testing continued...

Page 12: ...ish Contact Details Manufacturer Buckleys UVRAL Ltd Address Buckleys House Unit G Concept Court Shearway Business Park Shearway Road Folkestone Kent CT19 4RG UK Tel 44 0 1303 278888 Email sales buckle...

Reviews: