BÜCHI Labortechnik AG
Operation | 6
Operation Manual Pure Chromatography Instrument C-805
39/70
Add additional lines to the program collection
u
Tap the number field (1►).
ð
The display shows a drop-down list with selectable actions.
u
Select if the line should be added above or below the selected line.
ð
A line is added.
Deleting lines from the program collection
u
Tap the number field (1►).
ð
The display shows a drop-down list with selectable actions.
u
Select delete.
ð
The line is deleted.
6.2.7
Editing the vapor sensor sensitivity
The vapor sensor detects solvent concentration in the ambient air.
The following sensitivity limits are available:
Sensitivity
Meaning
High
Used for non-volatile solvents
Medium
Compromise between the low and high setting
Low
Used for volatile or semi-volatile solvents
Off
The vapor sensor is off
Navigation path
➔
Tools
➔
Vapor Sensor and Limits
u
Navigate to the
Vapor Sensors
dialog according to the navigation path.
ð
The display shows the dialog box
Vapor Sensors
.
u
Tap the drop-down list next
[Vapor alarm sensitivity]
.
u
Select the sensitivity value you wish to use.
u
Tap the button
[Close]
.
ð
The dialog box closes.