Setting up the network connection between computer and NIRMaster™

In this step the communication connection between the NIRMaster™ and computer is set up.

u

Start the computer, and open the dialog box 

 Ethernet Properties

.

u

Highlight the option 

[Internet Protocol Version 4 (TCP/IPv4)]

 from the

list.

u

Click the button 

[Properties]

.

ð

The dialog box 

Internet Protocol Version 4 (TCP/IPv4) 

open.

u

Activate the radio button next to 

[Use the following IP address]

.

u

Enter the following values.

• IP address: 192.168.1.2

• Subnet mask: 255.255.255.0

u

Click the button 

[OK]

.

ð

The dialog box 

Internet Protocol Version 4 (TCP/IPv4)

 close.

u

Click the button 

[OK]

.

ð

The dialog box 

Ethernet Properties

 close.

The network settings are now correctly set up so communication between the NIRMaster™ and computer is possible.

Start NIRWare operator software and verify that the Computer is correctly connected to the NIRMaster™.

NIRMaster™ Supplement Installation Manual