7 | Cleaning and servicing
Büchi Labortechnik AG
36/46
Operation Manual NIRMaster™
Fuse ok
Fuse triggered
Precondition:
R
The primary fuses are located inside the instrument next to the interferometer
power supply socket.
u
Switch off the instrument.
u
To get access to the resettable fuses, lift off the top cover as described in
Chapter 7.3 "How to lift off the top cover", page 31.
u
Push in the fuse for reset.
u
Switch on the instrument.
7.10
Exchanging the lamp
CAUTION
Risk of minor burns by hot lamp module.
u
Do not touch hot parts or surfaces.
u
Let the lamp module cool down.
u
Do not touch the bulb glass.
NOTE
After lamp-module exchange, a reference spectrum should be generated to match
the individual light intensity output of the module. See Chapter 6.3 "Reference
measurements", page 25.