Tightening torque: 2 Nm ± 0.5
u
Fix the ground cable to the sensor.
5.8
Connecting the video cable (optional accessory)
Tools required
Torque wrench, size 7 mm AF
u
Slide the cable guard over the video ca-
ble connector.
u
Plug the connector into the sensor.
Tightening torque: 2.5 Nm ± 0.5
u
Fix the cable guard to the sensor.