Working with the Autosampler K-71
8
K-370/371
Quick Guide, Version B
3
Working with the Autosampler K-371
The methods described in the following just show one of various possibilities. Other procedures are
described in the Operation Manual.
3 .1
Defining blanks using the express rack
Select
Configurator > Rack
and press
[New]
. Select the Group and User to which you want to
assign the new Rack (for Group definition see chapter 4.1.1).
Press
[Page]
twice and enter a name for the rack with the blanks.
Select
Express
by pressing the Tab key
twice.
Press the Enter key .
Press the Tab key again.
Press the Enter key .
Enter the information for Blank 1 and press
[Save]
and then
[Next]
.
Enter the information for Blank 2 and press
[Save]
and then
[Next]
.
Enter the information for Blank and press
[Save]
.
Press
[Close]
.
3 .2
Defining a sample rack and adding sample information to this rack
Select
Configurator > Rack
and press
[New]
. Select the Group and User to which you want to
assign the new Rack.
Press
[Page]
twice and enter a rack name.
Press
[Save]
. Press the Tab key and then press the Enter key .
Enter the sample information for Sample 1, then press
[Save]
.
Press
[Next]
and enter the sample information for Sample 2.
Continue until you have entered all samples.
Press
[Save]
, then press
[Close]
.
3 .3
Measuring blanks
Place the blanks for the measurement into the Express rack.
Select
Operator > Determination > Rack with sampler
and search for the rack you have defined
with the blank samples.
Press the Enter key .
Press
[Run]
to start the measurement of the blanks.
3 .4
Blank consideration for samples in a rack
Select
Operator > Determination > Rack with sampler
. Search for the Rack with the samples
for which you like to use the blanks and press the Enter key , then press
[Blank]
. Select
[Assign
blanks manually]
, press the Enter key .