BÜCHI Labortechnik AG
Cleaning and servicing | 11
Operation Manual FatExtractor E-500
75/88
u
Push the solvent tank up.
u
Tilt the solvent tank at 20° and remove it.
11.3
Draining solvent from the instrument
1. Switch on the pump. See Chapter 11.3.1 "Switch on the pump", page 75
2. Open the tank valves. See Chapter 11.3.2 "Open the tank valves", page 75
3. Switch of the valves. See Chapter 11.3.3 "Switch of the pump", page 76
11.3.1
Switch on the pump
Navigation path
➔
➔
[Service]
➔
[Actors]
Precondition:
R
A tank bottle is installed. See Chapter 5.8 "Preparing the tank bottle", page 30
u
Navigate to
[Actors]
via the navigation path.
ð
The display shows the view
Actors
.
u
Enable the action
[Pump]
.
11.3.2
Open the tank valves
Navigation path
➔
➔
[Service]
➔
[Actors]
➔
[Valves]
u
Navigate to
[Valves]
via the navigation path.
ð
The display shows the view
Valves
.
u
Select the position of the tank valve you wish to open.
u
Open the tank valve.
u
Wait until all solvent is in the tank bottle.