background image

20                                                                                                                                                                                                                           

6

7

8

6.Open the test strip vial. Take one 

new test strip out of the vial and 

recap the vial quickly and firmly.

7.Insert the test strip into the meter 

while the meter is off. The meter will 

turn on automatically. A blinking 

blood icon will be displayed on the 

screen. If no action is taken in 2 

minutes, the meter will turn off 

automatically.

8.Use the adjustable lancing device 

or the disposable lancet to take a 

blood sample from the fingertip and 

let it form a small round blood drop.

To avoid body fluid contamination, 

it is recommended that you do not 

use the first blood drop for glucose

testing.

Summary of Contents for 5228B

Page 1: ...5228BT M01A1 Rev date 11 2013 USER GUIDE Model 5228B...

Page 2: ...r to safety related notes in the manual accompanying this instrument Consult instructions for use Keep away from sunlight Authorised representive in the european community Manufacturer Do not re use U...

Page 3: ...Blood Glucose Test 11 Recalling the Memory and Viewing the Average 12 Expected Values 13 Transmitting Results Maintenance 14 Maintenance of Your System 15 Troubleshooting 16 Customer Service 17 Limita...

Page 4: ...us Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus nor for use with neonates The alternative site testing forearm in this system can only be use...

Page 5: ...ancets 8 Lancing Device 9 Package Inserts for Test Strips Control Solution and Lancets Lancing Device A single level control solution Medium is provided in the Starter Kit and no control solution is i...

Page 6: ...2 LCD Display Guide you through the test using symbols and simple messages 3 M Key Power ON OFF also for memory recalling mode please refer to manual for detailed function description 4 C Key Setting...

Page 7: ...ip When to use AST Medication stress illness food and exercise can affect blood glucose levels Capillary whole blood at the fingertip can reflect test changes faster than capillary blood at other site...

Page 8: ...or you to get a drop of blood for AST 2 Insert a new lancet firmly into the lancet holder 3 Twist off the protective top of the lancet 4 Close the end cap of the lancing device and grasp the lancing d...

Page 9: ...of the vial and quickly recap the vial to keep air out 8 Insert the test strip into the meter while the meter is off The meter will turn on automatically A blinking blood icon will be displayed on the...

Page 10: ...tting sufficient blood for a test it may be helpful to rub the puncture site a little longer before using the lancing device 11 Apply your blood to the front edge the tip of the test strip and the met...

Page 11: ...icapped persons near you when you conduct a glucose test using this system Please keep the test strip vial away from children The test strips and vial cap can present a choking hazard Please be cautio...

Page 12: ...inger and let a blood drop form 3 Apply the blood drop to the front edge of the test strip when the blood drop icon is still blinking on the meter Wait for 5 seconds and the meter will display the tes...

Page 13: ...er in the opposite direction of the arrow to close the cover into position 11 Note 1 It is suggested that batteries need to be replaced when the battery icon appears empty If you keep using the meter...

Page 14: ...ution test flag c 5 Indicates the meter is ready to take the blood sample when it flashes 6 Indicates the battery status 7 Indicates current displayed result is an average 8 Appears when alarm is on 9...

Page 15: ...2029 After the year setting is confirmed you will see the month segment flashing on the screen Set the Month Press C key to adjust the month until the desired month is displayed and then press the M k...

Page 16: ...clock in the lower left corner with a time segment display it means that you are in the alarm setting mode If you see only AL without the icon of the small alarm clock it means that you are not in the...

Page 17: ...mmol L is the standard unit of measurement in the Europe If the meter is to be sold in the Europe the unit of measurement will be set to mmol L by default Delete Memory Press C key to toggle between a...

Page 18: ...esults obtained during the 7 14 28 60 and 90 calendar days preceding the current date and time settings Note Your meter can show results either in mg dL or mmol L units It is very important to use the...

Page 19: ...of screen indicating the control test mode The test result will not be stored in the memory 3 Squeeze a small amount of Control Solution on a flat surface and apply to the front edge of the test strip...

Page 20: ...vial discard the first drop of control solution and wipe off the dispenser tip to ensure a good sample and an accurate result 6 Use only for 3 months after first opening Record the open date on the C...

Page 21: ...he end cap counter clockwise 2 Insert a new lancet firmly into the lancet holder 3 Twist off the protective tip of the lancet 4 Close the end cap of the lancing device as shown Slide into locking posi...

Page 22: ...automatically A blinking blood icon will be displayed on the screen If no action is taken in 2 minutes the meter will turn off automatically 8 Use the adjustable lancing device or the disposable lanc...

Page 23: ...nd or by pushing the ejector button The meter will turn off and the test result is stored auto matically Important 1 Never reuse a test strip or a lancet An E_5 error message will be displayed in the...

Page 24: ...f press M key to turn on the meter After a brief full screen display the meter will display the total number of test data stored in memory Then the screen will display 001 as the sequential number of...

Page 25: ...rward from 7 day average to 90 day average The average display will move from 90 days backward to 7 days if you continue to press C key 3 Press the M key and hold for only 2 seconds to exit the recall...

Page 26: ...ues 1 Reference SuperCheck Plus meter gives plasma equivalent results Before eating 5 6 mmol L 100mg dL Two hours after meals 7 8 mmol L 140mg dL Blood glucose levels normally will vary from time to t...

Page 27: ...he meter has a shelf life of 2 years 14 Maintenance of your system We suggest you should periodically compare the test system to another test system which is well maintained and monitored by a healthc...

Page 28: ...ple is not sufficient HI The glucose level is too high LO The glucose level is too low Following is a summary of all Error Messages These messages help to identify certain problems but do not appear i...

Page 29: ...lease call your medical doctor for advice immediately The meter is operating in an ambient temperature below 10 or 50 which is the lowest range of operating temperature 1 The meter displays an E_1 err...

Page 30: ...please contact Biotest Medical Corporation No 3 2 Chien kuo Road TEPZ Tantzu Taichung 427 Taiwan R O C TEL 886 4 2532 6668 E mail info mail biotestsystems com For questions related to your health con...

Page 31: ...ructose observed in blood glucose test 7 Use only Heparin or EDTA for anticoagulation of fresh capillary or venous whole blood 8 Xylose Do not test blood glucose during or soon after xylose absorption...

Page 32: ...ccuracy of Blood Glucose Monitoring System was SuperCheck Plus assessed by comparing blood results obtained by patients with those obtained using a YSI Model 2300 Glucose Analyzer which is calibrated...

Page 33: ...edical Corporation No 3 2 Chien kuo Rd TEPZ Tantzu Taichung 427 Taiwan R O C E mail info mail biotestsystems com Reference 1 American Diabetes Association Diabetes Care January 2007 volume 30 Suppl 1...

Page 34: ...and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help FCC Caution Any...

Page 35: ...BiotesT Medical Corporation No 3 2 Chien kuo Road TEPZ Tantzu 427 Taichung Taiwan ROC Tel 886 4 25326668 Fax 886 4 25326528 MEDNET GmbH Borkstrasse 10 48163 M nster Germany 0537...

Reviews: