V. Measuring
5.12.1.1 Setting of standard methods
Method 900:
Method 901:
5.12.1.2 Method Mode
• Mode: "fixed", fixed increment titration.
Depending on the time increment indicated, the Mütek
TM
PCD-05 records measured values at fixed intervals.
Addition amount has to be preset dynamic or conti-
nuous in the addition screen see chapter 5.12.1.3, page
46.
Mode: "adapt", adaptive titration.
As soon as an equilibrium is reached within a time
window, the Mütek
TM
PCD-05 continues with the next
dispensing step.
dE [mV]/ dt [s]:
Time window (dt) in which the variation of the mV signal
is below dE.
t(min) [s]/ t(max) [s]:
This display gives the minimum and maximum wait
times between two dispensing steps.
Addition amount has to be preset dynamic or continuos
in the addition screen see chapter 5.12.1.3, page 46.
Termination
mV
EP
0 mV
Pump
Automatic
Type
EP
Meas. Mode
fixed
Increment
3 s
Addition
Dynamic
dE (set)
8 mV
dV (min)
0,01 ml
dV (max)
0,1 ml
Termination
pH
EP
7,00 pH
Pump
Automatic
Type
IEP
Meas. Mode
Adaptive
dE
1,0 mV
dt
1 s
t (min)
5 s
t (max)
15 s
Addition
Dynamic
dE (set)
8 mV
dV (min)
0,005 ml
dV (max)
0,1 ml