ADMIN Menu
6.2.5 AZT
This stands for Automatic Zero Tracking. The
Div
value defines a ± range around zero.
When scale weight is not at the center of zero but inside this range for the time value
entered, ½ of the weight will be subtracted. This process is repeated until weight is
inside the center of zero region.
Div
Set the AZT window size to define stability. 3 divisions is the default value.
Set
Div
to
0
to disable AZT.
Time
Set the time window in seconds. 1 second is the default value. Set to
0
to
disable AZT.
6.2.6 Filter
Use this to filter out vibrations affecting the scale. Under this item you have the
following three parameters to set.
Avg
This stands for average. 10 is the default value.
Const
This stands for constant. 1 is the default value.
T-hold
This stands for threshold. 100 is the default value. When
0
is the threshold
value, filtering is always on.
To find the best settings for your filter needs, follow steps
1.
Using the Threshold setting is only recommended if the items to be weighed are
similar in weight. If the item weights vary considerably then set the threshold
value to 0 and proceed to step 2.
To determine the threshold value first set
T-Hold
to 0.0,
Const
to 0, and
Avg
to
1.0. Return to weigh mode and, with a typical item on the scale, observe the
weight swings. Record the difference between the highest and lowest displayed
weight values. Add 30 to 50% to this value. This is a good starting value for the
T-Hold s
etting. Do not set your indicator to this value until told to in step
.
2.
Setting
Avg
to higher values increases the filtering effect.
To do this: Set
T-Hold t
o 0.0,
Const
to 0, and
Avg
to 10. Check the stability of
the scale by exiting to normal weigh mode, remove all weight from the scale
and/or press
ENTER
and observe the Center of Zero annunciator. If it is on all
the time your scale is stable. If the Center of Zero light blinks, more filtering is
required. Go to step
3
.
Default settings are:
AVG = 10
Const = 1
t-hold = 100 (factory default based on 5000 lb capacity)
These values will provide the best weight response for the majority of scale
installations. In adverse conditions, where wind, vibration or other conditions are
affecting the stability of the weight displayed, refer to the following instructions to
improve the performance of the indicator.
Summary of Contents for T419S
Page 2: ...BTEK_T419S_s_en_501788 book 2019 B TEK All rights reserved ...
Page 118: ...118 T419S Indicator Service Manual ...
Page 130: ...130 T419S Indicator Service Manual ...
Page 134: ...134 T419S Indicator Service Manual Last 28 Temp 28 ...
Page 135: ......
Page 136: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088 ...