background image

3 1/2 DIGIT DIGITAL MULTIMETER WITH   

TRANSISTOR, FREQUENCY AND CAPACITANCE   

OPERATION MANUAL 

1.  SAFETY RULES 

 

This  meter  is  designed  and  tested  in  accordance  with  EN  publication  61010-1, 
installation category II overvoltage 600V. and pollution degree 2. 

 

This meter has been tested according to the following EMC Directives 

 

89/336/EEC (EMC of Nov., 1992, Electromagnetic Compatibility)

 

 

73/23/EEC Product safety law of Low Voltage Directive, EN61010-1 (04,93) 

 

This meter is designed to be indoor use at temperature 5

C to 40

C and altitude up to 

2,000m. 

 

To  ensure  that  the  meter  is  used  safely,  please  follow  all  safety  and  operating 
instructions  in  this  operation  manual.    If  the  meter  is  not  used  as  described  in  this 
operation manual, the safety features of this meter might be impaired. 

 

Do  not  use  the  meter  when  the  meter  or  test  lead  looks  damaged,  or  when  you 
suspect that the meter is not operating properly. 

 

When using the probes, keep your fingers behind the finger guards on the probes. 

 

Disconnect the live test lead before disconnecting the common test lead. 

 

Make  sure  power  is  off  before  cutting,  unsoldering,  or  breaking  the  circuit.    Small 
amount of current can be dangerous. 

 

Do not apply more than 600Vdc or 600Vac rms between a terminal and earth ground. 

 

To  avoid  electrical  shock,  use  CAUTION  when  working  above  60Vdc  or  25Vac  rms.   
Such voltages pose a shock hazard. 

 

Never make measurements with the battery cover or bottom case off. 

 

To avoid electrical shock or damage to the meter, do not exceed the input limits. 

 
2. 

INTERNATIONAL SYMBOLS

 

 

Important information 

 

Diode 

 

see manual 

 

Continuity 

 

AC 

 

Ground 

 

DC 

 

Double insulation 

 

3.  SPECIFICATION 

3.1 General Specifications 

Display 

:  3 1/2 digit LCD with maximum reading of 1999. 

Polarity 

:  Automatic, (-) negative polarity indication. 

Zero adjustment 

:  Automatic. 

Over range indication 

Only the massage “1” is displayed. 

Power 

:  Single, standard 9 volt battery NEDA 1604, JIS 006P, 

IEC6F22. 

Auto Power Off 

:  Approximate 15 minutes. 

(Pressing twice the Power 

Key to resume).

 

Dimension 

:  94 (W) x 205 (H) x 62 (D) mm. 

Weight 

:  Approx. 620

 

g (including battery and packaging). 

 
 
3.2 Electrical Specifications 
Accuracies are 

 (% of r no. of least significant digits) at 23

 5

C, <75% RH. 

 

DC Voltage 

Range  Resolution 

Accuracy 

Input Impedance 

Overload 

Protection 

200mV 

0.1mV 

(0.8% + 1) 

10M

 

250V DC/AC rms 

2V 

0.001V 

600V DC/AC rms 

20V 

0.01V 

200V 

0.1V 

600V 

1V 

AC Voltage 

Range  Resolution 

Accuracy 

Freq. Response 

Input Impedance   

10M

   

2V 

0.001V 

(0.8% + 3) 

40-400Hz 

Overload 

  Protection 

600V DC/AC rms 

20V 

0.01V 

200V 

0.1V 

600V 

1V 

(1.2% + 3) 

On 2V range, if short nor open the two test leads, display reading is 10 digits, this 
10 digits should not be subtracted from measurement results. 

DC Current 

Range  Resolution 

Accuracy 

Volt. drop 

Overload 

Protection 

Fast 200mA/250V 

Fused 

2mA 

0.001mA 

(0.8% + 3) 

200mV 

20mA 

0.01mA 

200mA 

0.1mA 

(1.0% + 3) 

20A 

0.01A 

(2.0% + 5) 

15sec. Max. Unfused 

AC Current 

Range  Resolution 

Accuracy 

Volt. 
drop 

Freq. 

Response 

Overload Protection 

20mA 

0.01mA 

(1.2% + 3) 

200mV  40-400Hz 

Fast 200mA/250V 

Fused 

200mA 

0.1mA 

(1.5% + 3) 

20A 

0.01A 

(3.0% + 5) 

15sec. Max. Unfused 

Resistance 

Range  Resolution 

Accuracy 

Open circuit volt. 

 

200

 

0.1

 

 

(0.8% + 2) 

 

<0.7V 

Overload 

Protection 

250VDC/AC rms 

<30sec 

2k

 

0.001k

 

20k

 

0.01k

 

200k

 

0.1k

 

2M

 

0.001M

 

20M

 

0.01M

 

(1.0% + 2) 

200M

0.1M

 

(5.0% + 10) 

<3V 

On 200M

Ω

  range, if short the two test leads, display reading is 10 digits, this 10 

digits should be subtracted from measurement results. 

Capacitance 

Range  Resolution 

Accuracy 

Test Freq. 

2nF 

0.001nF 

(2.5% + 5) 

400Hz 

20nF 

0.01nF 

200nF 

0.1nF 

2

0.001

20

0.01

On 2nF range, if Cx socket without DUT, display reading is 10 digits, this 10 digits 
should not be subtracted from measurement results. 

Frequency 

Range  Resolution 

Accuracy 

Overload Protection   

250VDC/AC rms <30sec 

2kHz 

0.001kHz 

(1.5% + 5) 

Vmin > 150mV 

Vmax < 10V 

20kHz 

0.01kHz 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Test voltage 

Test current 

Open circuit Volt. 

Overload 

Protection 

250VDC/AC rms 

<30sec 

Approx. 2.8V 

Approx. 1mA 

2.8V 

 

Range 

Test current 

Open circuit Volt. 

Buzzer sounds when 

resistance value   

≤ 60

 

Approx. 

1.0mA 

2.8V 

Transistor 

Test 

Range 

hFE    value 

Test volt.(Vce) 

Base current 

hFE 

0-1000 

Approx. 2.8V 

Approx. 10μA 

 
 
 
 
 
 
4. PANEL DESCRIPTIONS

 

 

1) 3 1/2 digits LCD display 
2) Power Key 
3) Function selector 
4) Rubber Holster 
5) Hold Key 
6) Transistor socket 
7) Input and common socket 
8) Capacitor socket 
 
   
 

 

 

5. 

OPERATION 

WARNING

 

1)  When  measuring  voltage  ensure  that  the  instrument  is  not  connected  or  switched  to  a 

current  or  resistance  or  temperature  or  to  the  diode  check/continuity  range.  Always 
ensure that the correct terminals are used for the type of measurement to be made. 

2)  Use  extreme  care  when  measuring  voltage  above  50V,  especially  from  sources  where 

high energy exists. 

3) 

Avoid making connections to “live” circuits whenever possible. 

4)  When  conducting  current  measureme

nts  ensure  that  the  circuit  is  not  “live”  before 

opening it in order to connect the test leads. 

5)  Before  conducting  resistance  measurements  or  continuity  /  diode  test,  ensure  that  the 

circuit under test is de-energised. 

6)  Always  ensure  that  the  correct  function  and  range  is  selected.  If  in  doubt  about  the 

correct range, starts with the highest and works downwards. 

7)  Extreme  care  should  be  taken  when  using  the  instrument  to  conjunction  with  a  current 

transformer connected to the terminals. High voltage may be produced at the terminals if 
an open circuit occurs. 

8)  Ensure  that  the  test  leads  and  probes  are  in  good  condition  with  no  damage  to  the 

insulation. 

9)  Take care not to exceed the over-load limits as given in the specifications. 
10) Fuse for replacement must be of the correct type and rating. 

 

5.1 DC and AC Voltage measurement 

1) 

Connect the black test lead to the “

COM

” socket and red test lead to the “

V

kHz

” socket. 

2) 

Set the function selector to “

V

” or “

V

” position to measure DCV or ACV. 

3)  Connect the test leads across the source or loads under measurement. 
 
5.2 DC Current measurement 
1) 

Connect the black test lead to the “

COM

” socket and red test lead to the “

mA

” socket for 

measurement up to 200mA. 

2) 

Set the function selector to desired resistance range “

A

” position. 

3)  Connect the test leads in series with the current source to be measured. 
4)  For  current  measurement  from  200mA  to  20A  (Unfused)  follow  generally  the  above 

procedure but connect the red test lead to “

20A

” socket. 

 
5.3 AC Current measurement 
1) 

Connect the black test lead to the “

COM

” socket and red test lead to the “

mA

” socket for 

measurement up to 200mA. 

2)  Set the function selector to desired resistance rang

e “

A

” position. 

3)  Connect the test leads in series with the current source to be measured. 
4)  For  current  measurement  from  200mA  to  20A  (Unfused)  follow  generally  the  above 

procedure but connect the red test lead to “

20A

” socket. 

 
5.4 Resistance measurement 

1) 

Connect the black test lead to the “

COM

” socket and red test lead to the “

V

kHz

” socket. 

2)  Set the function selector to desired resistance range 

 position 

3)  Connect the leads across the circuit to be tested. 

CAUTION: Ensure that the circuit to be tested is “dead”. 

                  Maximum input over-load : 250V rms < 30sec. 

 
5.5 Diode test 

1) 

Connect the black test lead to the “

COM

” socket and red test lead to the “

V

kHz

” socket. 

2)  Set the function selector to 

  (same as 

 

) position. 

3)  Connect the black and red test leads to the cathode (-) and anode (+) ends of the diode 

to be tested respectively. 

4)  Read the forward voltage drop (junction) value from the display. If reverse connected the 

test leads to diode, display shows over-load. 

 
5.6 Continuity test 

1) 

Connect the black test lead to the “

COM

” socket and red test lead to the “

V

kHz

” socket. 

2)  Set the function selector to 

  (same as 

) position. 

3)  Connect the leads across the circuit to be tested, if the resistance less than  or equal to 

approx. 60

, buzzer will be activated. 

CAUTION: Ensure that the circuit to be tested is “dead”. 

                 

    Maximum input over-load : 250V rms < 30sec. 

 

5.7 Transistor hFE test 
1) Set the function selector to 

hFE

” position. 

2) Determine whether the transistor is of NPN or PNP type and locate the E, B and C leads. 

insert the leads into the proper holes in the socket 

N

P

N

P

N

P

 

on the front panel. 

3) The display will read the approximate hFE value at the test condition of base current 10

μ

!

!

!

1

2

3

4

5

6

7

8

NPN
PNP

600V

MAX

Cx

20A

COM

mA

V

200mA MAX

FUSED

20A 15 sec. MAX

UNFUSED

MAX CAT II 

600V

kHz

Hz

20k

2k

20 F

20nF

2 F

200nF

2nF

POWER

HOLD

2V

20V

200V

600V

600V

20A 200mA 20mA

20A

2mA

20mA

200mA

A

A

V

200k

20k

2M

2k

20M

200

200M

hFE

20V

2V

200mV

200V

V

AUTO POWER OFF

!