background image

 

 

1.4 

Operational Procedure 

 

After switching on the units, allow one minute before the taking of readings on the detector, 

as the internal circuitry must be allowed to stabilise before correct results will be indicated.  

When testing any type of equipment, always check for system operation by operating the 

phase  reverse  switch  on  the  source.    A  corresponding  reversal  will  be  indicated  by  the 

phase  LED's  on  the  detector.    This  will  serve  as  a  check  that  the  system  is  functioning 

correctly. 

 

1.5 

Signal Levels 

 

Always ensure the source's output level is sensible for the equipment under test, and does 

not  cause  overloading.    Excessive  hum  or  noise  may  cause  indeterminate  results,  with 

detector LED's alternately lighting. 

 

1.6 

Source Frequency 

 

This must be set appropriately for the equipment under test, especially where it may have 

frequency dependant parts and contain major phase shifts.  Refer to the following sections 

for  specific  details.    When  supplied,  the  frequency  ranges  will  have  been  set  to  LOW  = 

110hz;  

M1 = 500Hz; M2 = 3k5Hz; H = 10kHz. 

 

2.0 

APPLICATIONS 

 
2.1 

Passive Components 

 

This includes connecting cables, transformers, line attenuators, etc.  Connect the source to 

the input, and the detector to the output, observing all the points as in section one above.  

The source output level switch can be at OdB, and the tone frequency set to band Ration 

M1.  High ratio transformers should be tested in a step-down mode. 

 

2.2 

General Equipment 

 

This  includes  all  forms  of  signal  processing  units.    Those  containing  filters  or  other 

frequency/phase  changing  parts,  should  have  their  cancel  or  bypass  switches  operated.  

Connect the source to the input, and the detector to the output, observing all points as in 

section one above.  The source output level should be set to suit the particular unit, to avoid 

any overload within it.  The detector has a wide dynamic range for its input level and will 

accept  anything  from  a  few  millivolts,  up  to  fifty  volts.    For  all  wide-band  equipment,  the 

source tone frequency can be set to either range M1 or M2. 

 

2.3 

Microphones 

 

For phantom powered types, a separate powering unit will be required.  For high impedance 

moving  coil  types,  a  step-down  transformer  will  be  required.    Connect  the  source  to  a 

loudspeaker,  either  directly  or  via  a  power  amplifier,  and  adjust  to  give  a  comfortable 

listening  level.    The  tone  frequency  can  be  set  to  either  range  M1  or  M2.    Connect  a 

microphone with a known phase wiring to the detector, and hold up to the loudspeaker, on 

axis,  and  observe  the  phase  LED  colour.    Repeat  with  the  unknown  microphone,  and 

compare  phase  LED  colour  with  that  of  the  known  reference.    For  some  microphones  it 

might  be  necessary  to  adjust  the  tone  frequency  to  avoid  a  particular  peak  or  dip  in  the 

phase response. 

 

 

www.audiovias.com

Reviews: