21
AC current (20000-count mode)
RANGE
Accuracy
Burden Voltage
Model 839
Model 836
40Hz ~ 3kHz
199.99
A, 1999.9
A
1% + 20d
1.2% + 20d
0.2mV/
A
19.999mA, 199.99mA
3mV/mA
1.9999A, 10.000A
1)
1% + 40d
1.2% + 40d
30mV/A
3kHz ~ 5kHz
199.99
A, 1999.9
A
2% + 20d
Unspecified
0.2mV/
A
19.999mA, 199.99mA
3mV/mA
1.9999A, 10.000A
1)
Unspecified
30mV/A
1)
10A continuous up to ambient 35
o
C; <15 mins on per >5 mins off @ 35
o
C ~ 50
o
C.
>10A to 20A for <30 seconds on per >5 mins off
~ Hz Line Level Frequency
Function RANGE
Sensitivity (Sine RMS)
Range
199.99mV, 600.0mV
40mV
6Hz ~ 100kHz
1.9999V, 6.000V
0.4V
10Hz ~ 50kHz
19.999V, 60.00V
4V
199.99V, 600.0V
40V
10Hz ~ 30kHz
1000.0V, 1000V
400V
10Hz ~ 5kHz
VFD 600.0V
40V
10Hz ~ 440Hz
VFD 1000V
400V
10Hz ~ 440Hz
199.99
A, 600.0
A
40
A
10Hz ~ 5kHz
1999.9
A, 6000
A
400
A
19.999mA, 60.00mA
4mA
199.99mA, 600.0mA
40mA
1.9999A, 6.000A
0.6A
10Hz ~ 3kHz
10.000A, 10.000A
6A
Accuracy: 0.01%+4d