11
6) SPECIFICATIONS
Display : 3-3/4 digits 4000 counts LCD display(s)
Update Rate : 3 per second nominal
Polarity : Automatic
Low Battery : Below approx. 2.4V
Operating Temperature : 0
o
C to 40
o
C
Relative Humidity : Maximum relative humidity 80% for temperature up to 31
o
C
decreasing linearly to 50% relative humidity at 40
o
C
Altitude : Operating below 2000m
Storage Temperature : -20
o
C to 60
o
C, < 80% R.H. (with battery removed)
Temperature Coefficient : nominal 0.15 x (specified accuracy)/
o
C @(0
o
C -18
o
C or 28
o
C
-40
o
C), or otherwise specified
Sensing : Average sensing for Models
061s & 063s
; True RMS for Models
062s & 065s
Safety : Meets
IEC/UL/EN61010-1 Ed. 3.0
,
IEC/EN61010-2-032 Ed. 3.0
,
IEC/EN61010-2-033 Ed. 1.0, IEC/UL/EN61010-031 Ed. 1.1 &
CAN/CSA-C22.2 No.
61010-1-12 Ed. 3.0
:
Category III 600 Volts ac & dc.
Transient protection : 6.5kV (1.2/50
s surge) for all models
Pollution degree : 2
E.M.C. : Meets EN61326-1:2006 (EN55022, EN61000-3-2, EN61000-3-3,
EN61000-4-2, EN61000-4-3, EN61000-4-4, , EN61000-4-5, EN61000-4-6,
EN61000-4-8, EN61000-4-11)
In an RF field of 3V/m:
Capacitance function is not specified
Other function ranges: Total Accuracy = Specified Ac 50igits
Performance above 3V/m is not specified
Overload Protections :
Clamp-on jaws : DC/AC 400A rms continuous
+ & COM terminals (all functions) : 600VDC/VAC rms
Power Supply : standard 1.5V AAA Size (NEDA 24G or IEC R03) battery X 2; or 1.5V
AAA Size (NEDA 24A or IEC LR03) alkaline battery X 2
Power Consumption : 11mA typical for ACA/DCA and 2.9mA typical for other functions
APO Timing : Idle for 30 minutes
APO Consumption : typical 10
A for Model
061s & 063s
and 190
A for Model
062s &
065s
Dimension : L188mm X W63mm X H40mm
Weight : 218 gm approx
Jaw opening & Conductor diameter : 30mm max