Operation
Z33119_1_003
33
7.3
Manual Operation Modes BCU I
Rotary switch
Operation
Notes
The BCU I is controlled by the BSVT (see
Remote Operation Modes on BSVT [
) or BVT (see
Remote Operation Modes on BVT [
Flush / 0 *
No cooling power
This setting is used when VT gas at room temperature has to be fed to the NMR probe or the
VT gas tubing of the BCU I has to be dried and purged. In this setting the compressor is not
running.
Purge with dry clean air or nitrogen gas at room temperature with a dew point temperature ≤
-50°C@1bar
1 *
Medium cooling power
In this mode, the VT gas is cooled only slightly with reference to room temperature T
RT
(T =
T
RT
0…20°C)
2 *
Maximum cooling power
In this maximum mode, the VT gas is cooled to the lowest specified temperature.
* ) The BCU I is in a manual mode and cannot be controlled with the BVT or BSVT. In manual mode the BCU I can be operated without a communication
cable to the B(S)VT.
7.4
Remote Operation Modes on BSVT
Rotary switch
TopSpin EDTE
Operation
Notes
Remote
Off
No cooling power
This setting is used when VT gas at room temperature has to be fed to the
NMR probe or the VT gas tubing of the BCU I has to be dried and purged.
In this setting the compressor is not running.
Purge with dry clean air or nitrogen gas at room temperature with a dew
point temperature ≤ -50°C@1bar
Remote
Medium
Medium cooling power
In this mode, the VT gas is cooled only slightly with reference to room
temperature T
RT
(T = T
RT
0…20°C)
Remote
Maximum
Maximum cooling power
In this maximum mode, the VT gas is cooled to the lowest specified
temperature.
Summary of Contents for SmartCooler BCU I
Page 1: ...SmartCooler BCU I User Manual Version 003 Innovation with Integrity NMR...
Page 12: ...Safety 12 Z33119_2_003...
Page 18: ...Introduction 18 Z33119_2_003...
Page 28: ...Preparation 28 Z33119_2_003...
Page 30: ...Installation 30 Z33119_2_003...
Page 42: ...Troubleshooting 42 Z33119_1_003...
Page 44: ...Replacement of Parts 44 Z33119_2_003...
Page 48: ...Contact 48 Z33119_2_003...
Page 49: ...Z33119_2_003 49...
Page 50: ...Bruker Corporation info bruker com www bruker com Order No Z33119...