60
Z31900_00_01
Operation Overview
8.2.3
Sample Adjustment and Maximum Sample Height
Before the spinner is inserted into the cassette:
1.
Adjust the spinner using the sample depth adjustment gauge.
2.
Insert the spinner at the
Operator Access Position
into the holder.
3.
Check if the sample (
/2) is under the maximum sample height.
The glass
tube must be below the yellow line, otherwise it will break
If the sample is too long,
do not use it
in the SampleXpress!
DANGER
Danger of injury due to glass tube breakage!
Broken glass tubes may cause minor to severe injuries as well as material damage.
Substances escaping the broken glass tubes may be hazardous or allergenic and
therefore cause minor or severe injuries or even death.
f
Wear protective equipment.
f
Refer to the corresponding precautions and cleaning/disinfection instructions.
f
Perform all tasks with the cassette and glass tubes carefully.
f
Never turn the cassette upside down or turn it on one side.
Figure 8.2
Operator Access Position
1. Maximum Sample Height
2. Sample
3. Spinner
Summary of Contents for SampleXpress
Page 11: ...11 Z31900_00_01 Introduction 2 6 EC Declaration of Conformity ...
Page 12: ...12 Z31900_00_01 Introduction 2 7 EG Konformitätserklärung ...
Page 56: ...56 Z31900_00_01 Installation and Initial Commissioning ...
Page 104: ...104 Z31900_00_01 Maintenance ...
Page 116: ...116 Z31900_00_01 Error Codes ...
Page 120: ...120 Z31900_00_01 Dismantling and Disposal ...
Page 122: ...122 Z31900_00_01 Contact ...
Page 144: ...144 Z31900_00_01 Avertissements De Sécurité ...
Page 156: ...156 Z31900_00_01 Sicherheitshinweise ...
Page 158: ......
Page 160: ...160 Z31900_00_01 Warning Signs ...
Page 164: ...164 Z31900_00_01 Figures ...
Page 166: ...166 Z31900_00_01 Tables ...
Page 168: ...168 Z31900_00_01 Glossary ...
Page 172: ...172 Z31900_00_01 Index ...
Page 173: ......