background image

Sample Insert (down) Problems

NMR Case
Trouble Shooting Guide 

BRUKER

Page 5 of 15

1.8

 

Interface tube inside the adapter collar

Inside the adapter collar is a tube, which interfaces the adapter collar with the BST shim upper stack (see Figure 2 and 3 at
the end of this manual and Figure 3 at the end of the installation manual). The correct seating of the adapter collar and the
interface tube is critical for the reliability of the sample changer.

To check the seating, first remove the 3 screws and then remove the NMR Case base plate, leaving only the collar in place.
Check that the bottom of the collar is flush with the BST. Check that the interface tube is flush with the top of the collar.
Check that the five fingers of the interface tube are not bent inwards and block the inside diameter of the BST.

See figure 4 at the end of the installation manual for the correct seating. See figure 5 at the end of the installation manual for
the incorrect seating.

If the collar cannot be seated correctly, then check the BST top head for parts sticking out (labels or covers).

If the O-ring does not engage onto the BST (it feels too tight), then apply some silicone grease onto the O-ring inside the
collar.

If the interface tube is not flush with the top of the collar, then it may be an old version (made from brown plastic). Replace it
with a new version made from black metal (Part number B3695).

If the inner diameter of the interface tube seems to be too small, then it may be an old version (made from brown plastic).
Replace it with a new version made from black aluminum metal (Part number B3695).

2.  Sample Insert (down) Problems

2.1

 

NMR Case or magnet not level

The shim stack must be vertical and the NMR Case horizontal aligned. Check it with the circular level (supplied with the tool
kit). The bubble must be inside the innermost ring. Adjust the magnet stand and the telescopic front feet as necessary.

2.2

 

BST shim stack with sharp edges

On top of the BST, where the samples are inserted, there are some cutouts located on the inside diameter. Some cutouts have
sharp edges. If this is the case (check it with a finger), either replace the BST or remove the sharp edges with a file.

2.3

 

Sample lift air bypass

After changing a sample, the new sample should stays on top of the sample holder tray for a short moment, before it is
rapidly inserted into the BST (within 2 seconds).
There is a small pneumatic cylinder located inside aluminum housing underneath the NMR Case base plate (see Figure 2 and
Figure 3). This cylinder opens and closes a sample lift air bypass exit hole, located in the adapter collar. Check for the correct
opening and closing of this bypass exit hole.
If the bypass exit hole is not open completely, the sample does not get inserted fast enough into the BST and may get stuck
on top of the BST.
If the bypass exit hole is not closed completely, the sample will not eject to the top of the sample tray, but will sit on top of
the BST instead.

Summary of Contents for NMR CASE

Page 1: ...BRUKER NMR CASE TM Trouble Shooting Guide ...

Page 2: ...ring installation or operating of the items or units described in this manual Unauthorized reproduction of manual contents without written permission from the publishers or translation into another language either in full or in part is forbidden Written by Martin Rindlisbacher BRUKER BioSpin Corporation 15 Fortune Drive Billerica MA 01821 USA martin rindlisbacher Bruker Biospin com Version 10 July...

Page 3: ...rmittent problems 10 11 Sample lift eject sometimes hangs on top of the BST and does NOT go up into the NMR Case 11 12 Sample lift eject always hangs on top of the BST and does NOT go up into the NMR Case 11 13 Sample insert down sometimes hangs on top of the sample holder tray 11 14 Sample insert down sometimes hangs at the white ring on top of the adapter collar 12 15 The sample holder tray does...

Page 4: ...to disengage the light barrier Read the instruction sheet P N B2300 1 4 Seals To avoid sample lift air leaks the following four seals must be properly installed see installation manual Figure 1 page 17 1 Inside the adapter collar one O Ring 2 On top of the adapter collar one flat rubber seal 3 Inside the large white round ring plate one flat rubber seal 4 One white or brown seal ring on top of the...

Page 5: ...rt number B3695 If the inner diameter of the interface tube seems to be too small then it may be an old version made from brown plastic Replace it with a new version made from black aluminum metal Part number B3695 2 Sample Insert down Problems 2 1 NMR Case or magnet not level The shim stack must be vertical and the NMR Case horizontal aligned Check it with the circular level supplied with the too...

Page 6: ...gn debris like sticky tape glue from tape or staplers Check that the 3 white round plates located on the NMR Case base plate are clean and free of foreign debris like sticky tape glue from tape or staplers parts Switch off the motion controller and push the black pin on the side to the inside out of the way Mount the sample holder tray and check for smooth rotation by hand It should not go easy bu...

Page 7: ...der tray rotates ON O 1 0 Sample goes down insert OFF I 0 2 3 1 Rotating mechanism resets OFF O 1 0 Air ON always ON ON O 2 Next sample is inserted ON O 3 No sample is inserted OFF O 3 Reset button depressed ON OFF I 4 O 3 3 1 4 Check that I 0 is turned ON when the sample is on top of the sample holder tray The sample holder spinner must be pressed up firmly against the sample stop head of the tra...

Page 8: ...n 1 6 4 1 5 Four seconds after switching on Green I RUN ERR 0 1 2 3 4 5 O green COM I O 0 1 2 3 4 1 6 If the LED O 3 is green then press the RESET switch The O 3 LED should go off 4 1 7 If there are no lights check the power cord and the fuse There is a spare fuse inside the fuse holder 250V 3A slow blow 4 1 8 If during the first two seconds some lights are missing or a red light is still on after...

Page 9: ... manifold see Installation manual Figure 3 Use a small screwdriver and turn the adjustment screw clockwise CW in by quarter turn increments Check for slamming and turn further in if necessary However the airflow should not be decreased too much or the sample holder tray will no longer advancing firmly See section 2 5 of this manual 6 The adapter collar cannot be removed from the BST upper stack Th...

Page 10: ...ent manual P N B2928 section 10 9 The sample spin rate goes up to 150 Hz Check the following to see if the problem goes away 9 1 1 Unplug the spin rate cable connector at the BST shim stack Check it and plug it back in 9 1 2 Remove the BSMS sample control modifier cable B1428 see instruction sheet B2300 9 1 3 Check the spin rate settings 10 Intermittent problems The following 4 items need to be pr...

Page 11: ... ring brown tube Section 1 6 5 sample lift airflow Section 1 1 12 Sample lift eject always hangs on top of the BST and does NOT go up into the NMR Case Item Check out Trouble Shooting Guide 1 sample lift airflow Section 1 1 2 correct closing of the sample lift air bypass Section 1 7 3 Motion controller has no power Section 4 4 Motion controller has no air pressure Section 4 13 Sample insert down s...

Page 12: ...n 2 2 5 NMR Case or magnet not leveled Section 1 8 15 The sample holder tray does not advance away from position 0 Cause a sample has been removed manually or the NMR Case has been used in manual mode Solution Press the RESET button on the side of the controller while the sample lift is OFF to clear the sample tracking memory Note pressing the RESET button while the sample lift is ON does not work...

Page 13: ...ide BRUKER Page 13 of 15 17 Figure 1 NMR Case Top View Extend Stop Post Cylinder B Sensor 1 5 Cyl A L in 1 Cyl A L out 2 Cyl B Extend 3 Cyl B Retract 4 Cylinder A Sensor 1 Cyl A L in Cyl A L out Cyl B Retract Cyl B Extend Retract Stop Post NMR Case Top view ...

Page 14: ...Guide BRUKER Page 14 of 15 18 Figure 2 NMR Case Side View Base cover Adapter Collar Sensor 1 Sample lift air bypass valve Opens and closes Cyl B Extend Cyl B Retract Small Cylinder Front leg NMR Case Side view Sensor 2 Interface tube inside adapter collar ...

Page 15: ...ure 3 NMR Case bottom view Adapter Collar Interface tube Small cylinder 5 Sensor 1 1 Cyl A L in 2 Cyl A L out 3 Cyl B Extend Sample lift air bypass valve opens and closes 4 Cyl B Retract 6 SL Sensor Sample lift air Air flow adjustment screws 7 Sensor 2 Front leg Front leg Pneumatic manifold ...

Reviews: