164
HYPERION User Manual
Bruker Optik GmbH
Specifications A
FPA detectors for HYPERION 3000
The FPA
1
detectors listed in the table below are multi-element detectors.
A.6
Heating / freezing sample stage THMS600 (Linkam)
1. FPA - Focal Plane Array detector
Detector
Number of
detector elements
Spectral range
Mode of cooling /
Nominal hold time
photovoltaic MCT
FPA detector
16,384 elements
(square arrangement of
the elements:
128 x 128)
4,500 - 900 cm
-1
with liquid N
2
typical hold time when in
use: 8 h
typical hold time when not in
use: 20 h
photovoltaic MCT
FPA detector
1,024 elements
(square arrangement of
the elements: 32 x 32)
4,500 - 900 cm
-1
with liquid N
2
typical hold time when in
operation: 8 h
typical hold time when not in
operation 20 h
photovoltaic MCT
FPA detector
4,096 elements
(square arrangement of
the elements: 64 x 64)
4,500 - 900 cm
-1
with liquid N
2
typical hold time when in
operation: 8 h
typical hold time when not in
operation 20 h
Parameter
Specification
Temperature range:
-196°C to 600°C
Heating rates:
between 0.1 (min.) and 130°C/min (max.)
Temperature stability:
< 0.1°C
Heating element:
Silver block (ensures high thermal conductivity and allows
for extremely sensitive temperature measurements)
Stage body dimensions:
137 mm x 92 mm x 22 mm
Stage is dimensioned for:
sample diameter: 22 mm (max.)
sample thickness: 1.5 mm (max.)
Positioning range:
16 mm in x-direction and 16 mm in y-direction
Sample area:
22 mm diameter within the stage unit
Required working distance:
> 4.5 mm
Spectral range:
visible (quartz window)
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...