This manual is intended to be a quick-start reference guide for Bruker service techni-
cians. It provides basic information for installing and setting up the Fourier 300 system.
Carefully read the information in this manual, as well as, the safety information
provided in the Service Manual before working on the device!
This manual provides only the basic information required for installation and set-up, for
complete information refer to the service manual.
1.2
Before You Begin
This guide assumes that the magnet has been evacuated, cooled down, and that all the
equipment for energizing that magnet is available on site.
Three test samples, two spinners and a sample gauge are shipped with the system.
•
3% CDCl3 in Chloroform (lineshape)
•
0.1% EB in CDCl3 (1H SINO)
•
ASTM (C13 SINO)
To shim the magnet and perform the ATPs test you will need to bring some additional
samples:
•
Urea in DMSO (Pulse Length adjustment)
•
100mg or 1g Cholac in CDCl3 (DEPT)
•
Cyclosporine in C6D6 (2D Examples)
1.3
Unpacking the Console
Do not attempt to lift the Fourier console out of the carton
, rather:
1. Slide the carton sleeve up and over the console.
2. Slide the console off the pallet.
The console has wheels and can be moved around freely using the handle on the back.
Retain the packaging in case the console has to be returned for repair/exchange.
Summary of Contents for Fourier 300
Page 1: ...Version Innovation with Integrity Fourier 001 NMR Fourier 300 Quick Start Installation Guide ...
Page 4: ...4 Z33048_00_01 Table of Contents ...
Page 18: ...18 Z33048_00_01 Main Differences Between Fourier and AVANCE ...
Page 20: ...20 Z33048_00_01 Figures ...
Page 21: ......
Page 22: ...Bruker Corporation info bruker com www bruker com Order No Z33048 ...