Power Supply Modules
16
Bruker
BSMS/2 Mainframe Version 001
Troubleshooting
3.3
1.
First check if one of the PSB LED’s is not lit. If so, look at the two fans above
the power supply modules. If one of them doesn’t run, that could be the cause
for an overheat protection of voltage regulator. In this case the fuse doesn’thave
to be blown.(For fan replacement see Figure 8)
2.
Switch the BSMS/2 of and disconect the mains power cable.
3.
Verify that the primary voltage selection switch on the power supply module is
in the correct position (see Installation on page 14).
4.
Open the cover plate and replace the fuse corresponding to the LED that dosn’t
light. (For further informations see Fues 3.4)
5.
Now connect the mains power cable, turn on the mains switch and see if the
LED lights.
6.
If the same fault occurs again plug out all modules in the front and back rack on
which the defective supply is used (see Figure5 and 6). Replace the fuse (Step 1
to 4). Switch on the mains and if still a LED dosn’t light the defect is locatated
on this PSB. Otherwise plug in the first module and switch on the mains switch.
Check the LED’s and continue with the next module until a module causes the
DC interruption. The defectiv part is located and has to be replaced by the
BRUKER Troubleshooter.
7.
Change a fan.
Figure 9: Fan replacement 1
Disconnect all cables on front side
of the BSMS/2 and pull it out of the
cabinet.
remove Line Module screws
solder out the two wires
Summary of Contents for BSMS/2
Page 1: ...BRUKER BSMS 2 BSMS 2 Mainframe Technical Manual Version 001 Bruker ...
Page 28: ...Central Processing Board CPU 28 Bruker BSMS 2 Mainframe Version 001 ...
Page 36: ...Buses and Wiring 36 Bruker BSMS 2 Mainframe Version 001 Chassis wiring 6 2 ...
Page 37: ...BSMS 2 Mainframe Version 001 Bruker 37 User Bus wiring 6 3 ...
Page 38: ...Buses and Wiring 38 Bruker BSMS 2 Mainframe Version 001 JU1 6 are placed placed ...
Page 39: ...BSMS 2 Mainframe Version 001 Bruker 39 ...
Page 40: ...Buses and Wiring 40 Bruker BSMS 2 Mainframe Version 001 ...
Page 41: ...BSMS 2 Mainframe Version 001 Bruker 41 ...
Page 42: ...Buses and Wiring 42 Bruker BSMS 2 Mainframe Version 001 ...
Page 47: ...Chapter BSMS 2 Mainframe Version 001 Bruker 47 Schemata 8 Power supply PSB1 place plan 8 1 ...
Page 48: ...Schemata 48 Bruker BSMS 2 Mainframe Version 001 Power Supply PSB2 place plan 8 2 ...
Page 49: ...BSMS 2 Mainframe Version 001 Bruker 49 Variant Scheme PSB1 PSB2 8 3 ...
Page 50: ...Schemata 50 Bruker BSMS 2 Mainframe Version 001 ...
Page 51: ...BSMS 2 Mainframe Version 001 Bruker 51 ...
Page 52: ...Schemata 52 Bruker BSMS 2 Mainframe Version 001 ...
Page 53: ...BSMS 2 Mainframe Version 001 Bruker 53 ...
Page 54: ...Schemata 54 Bruker BSMS 2 Mainframe Version 001 ...
Page 55: ...BSMS 2 Mainframe Version 001 Bruker 55 ...
Page 56: ...Schemata 56 Bruker BSMS 2 Mainframe Version 001 ...
Page 57: ...BSMS 2 Mainframe Version 001 Bruker 57 ...
Page 58: ...Schemata 58 Bruker BSMS 2 Mainframe Version 001 ...
Page 59: ...BSMS 2 Mainframe Version 001 Bruker 59 ...
Page 60: ...Schemata 60 Bruker BSMS 2 Mainframe Version 001 ...
Page 61: ...BSMS 2 Mainframe Version 001 Bruker 61 Lastpage ...