50
Z31823_00_05
Hardware Overview
8.2.4
UPS
Uninterruptable Power Supply (ECL < 01.00: P/N O10440;
.00: P/N 1802515)
•
Input: 208V, 200V*, 208V**, 220V, 230V (default), 240V
*100V and 200V are derated by 20%
**110V and 208V are derated by 10%
Voltage range at 100%: 180-276 Vac.
•
1000VA
•
Bridging time: 7 minutes at full output load (1000VA)
•
19“ rack mount 2HE
For more details refer to the separate UPS manual.
8.2.5
Pump Cabinet
The pump cabinet consists of:
•
2 vacuum pumps (P1 and P2) and 2 valves (V1 and V2), connected to the pumping
line.
•
A helium flow meter which measures the helium flow at the exhaust of both pumps.
The flow meter (P/N 87487) is connected to the BMPC/2 control unit via the CAN
bus.
•
A temperature sensor (PT100) to measure the temperature T5 inside the pump cab-
inet.
ECL
01
≥