Configuration (administrator)
In the He Measurement Check window you can activate the helium level monitoring.
During the monitoring you can examine whether the helium changes over a preset
period of time. This time period can be varied within a range of 1 and 7 days using the
field „Set Monitor Period to x days“. When the value of the helium level remains stable
for the selected time a warning will be given and a warning light will be activated. With
this a missing or incorrectly initialized helium level measurement can be recognized.
12.3.3.4
Pulse Tube Cooler
If your magnet system includes a Pulse Tube Cooler (PTC) it is recommended that this
be monitored. The PTC is not monitored directly, but by monitoring the heater power
(Heater Power 1), it can be determined whether the PTC is operating properly.
Normally the heater power (Heater Power 1) must be above the set value. If the heater
power is below this limit for more than the set time this usual indicates there is a prob-
lem.
In this case the system turns on the warning lamp and the software causes the "Info-Dia-
log" to popup.
To clear the warning and to switch off the warning lamp you must confirm the warning in
the "Info-Dialog".
Figure 12.20Tools - Options - Pulse Tube Cooler
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......