Planning Details
42
Site Planning Information BioSpec 70/30 USR_2_V020
Unit
Number
Size
(width x length x
height)
A
B
Electronic Cabinets
2
each 60 x 80 x 200 cm
90 cm
55 cm
High Power Gradient
Cabinet
1
60 x 90 x 200 cm
90 cm
45 cm
Cooling Unit (DLHE)
1
80 x 80 x 200 cm
90 cm
55 cm
Line Power Distributor
(wall assembly)
1
76 x 22 x 76 cm
85 cm
-/-
Table 6.2:
Measurements of objects in the technical area
Unit
Number
Size
(width x length x
height)
A
B
CRP Cooling Unit
1
75 x 79 x 98 cm
70 cm
30 cm
CRP Compressor
1
45 x 49 x 59 cm
30 cm
30 cm
CRP Helium Tank
1
50 liter tank
-/-
-/-
Table 6.3:
Measurements of the MRI CryoProbe™ units in the technical area
6.2
RF Shielded Room
The purpose of a RF shielded room (Faraday Cage) is to shield the inner room of the cage
from disturbing RF electromagnetic radiation. The fundamental prerequisite for this is that:
• the room is designed to be free of unfiltered power lines and uncontrolled electromagnet
radiation penetration.
• the room is earthed solely at the central earth point of the MR instrument and is otherwise
electrically insulated and thereby free of ground loops.
Summary of Contents for BioSpec 70/30 USR V2
Page 6: ...Contents vi Site Planning Information BioSpec 70 30 USR_2_V020...
Page 20: ...Safety 20 Site Planning Information BioSpec 70 30 USR_2_V020...
Page 88: ...Moving or Dismantling the MR Instrument 88 Site Planning Information BioSpec 70 30 USR_2_V020...
Page 98: ...Tables 98 Site Planning Information BioSpec 70 30 USR_2_V020...
Page 103: ...Site Planning Information BioSpec 70 30 USR_2_V020 103...