46
ID: 182463115
Planning details
Portions of the exhaust gas line
Portions of the exhaust gas line
Section of the exhaust vent
Figure 6.10: Section of the exhaust vent
Calculation sections of the exhaust pipe
and table
provide an example for calculating the pressure drop of an exhaust vent. The posi-
tion corresponds to the entire length from the magnet to the start of the respective segment. The
respective standard values for the pressure drop for a pipe diameter of 200 mm have been taken
from the table
Pressure drop in USR magnet cable elements: 47/40, 70/30, 94/20,
6.4
Floor construction
A suitable floor construction must take into account the following specifications:
▪
adequate load-bearing capacity and load distribution
▪
floor and building vibrations
▪
impact sound decoupling
▪
ESD tolerance
Segment
L [m]
D [mm]
Pos [m]
DP [mbar]
1
0.5
200
0
0.5 x 1.4 = 0.7
1a
90° bends
200
0.5
20.0
2
1.8
200
0.5
1.8 x 1.4 = 2.52
2a
90° bends
200
2.3
20.0
3
3.5
200
2.3
3.5 x 1.4 = 4.9
3a
90° bends
200
5.8
40
4
2.0
200
5.8
2 x 2.8 = 5.6
4a
90°bends
200
7.8
40
5
4.0
200
7.8
4 x 2.8 = 11.2
5a
90°bends
200
11.8
60
6
4.0
200
11.8
4 x 4.2 = 16.8
Total
181.7
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...