B-ACS 60/120 Command Implementation
7.1
Release: 19990701
A command always consists of two letters which in some cases are followed by a
space and a parameter. The space between the command and the parameter is
essential. A command must always be terminated with <CR>. If this syntax is not
respected, the sample changer gives the message: "Invalid Parameter" or "Invalid
Command".
The answer of any command is terminated with <CR><LF> or still is only
<CR><LF>.
The sample changer software commands have been divided into three modes:
1. Operation Mode
This is the actual sample changer operating mode in which complete sam-
ple exchange sequences can be run by a host computer.
2. Error Mode
This mode has only two commands: one to continue the program after an
error has been observed, and one to move to the defined neutral position
"HOME".
3. Diagnostic Mode
This mode offers the user a range of commands to test all the different
sample changer functions.
The different commands are described in detail in the following sections.
Operation Mode
7.1.1
The following list gives a detailed description of all operation commands in alpha-
betical order.
Summary of Contents for B-ACS 120
Page 1: ...BRUKER Version Bruker Automated Control Systems B ACS 60 120 Installation Manual 001...
Page 7: ...Installation Manual Version 001 BRUKER 7 167 1 Declaration of Conformity 1...
Page 10: ...10 167 BRUKER Installation Manual Version 001 Introduction...
Page 40: ...40 167 BRUKER Installation Manual Version 001 Operating Instructions...
Page 56: ...56 167 BRUKER Installation Manual Version 001 Special Tools...
Page 60: ...60 167 BRUKER Installation Manual Version 001 Technical Data...
Page 150: ...150 167 BRUKER Installation Manual Version 001...
Page 160: ...160 167 BRUKER Installation Manual Version 001 Figures...
Page 167: ...Installation Manual Version 001 BRUKER 167 167 Lastpage...