Installation Manual Version 001
BRUKER
13 (167)
4
Mounting
Instructions
4
Caution
4.1
Before mounting the Automatic Sample Changer, it is important to make sure that
the magnet is firmly anchored to its base in order to prevent the instrumentation
from tipping over. When doing so, carefully align the legs and lifting lugs used to
secure the Sample Changer. The Sample Changer is held to the magnet by four
adjustable arm assemblies (two for magnets with active dampers or TMC legs).
The arm assemblies should be aligned with the magnet as shown in
Figure 4.1. Location of Adjustable Arm Assemblies
LIGHT BARRIER
LIFTING LUGS
COLUMN
LEGS
ADJUSTABLE
ARM
ASSEMBLIES
SUPPORT FOR MAGNETS
WITH VIBRATION DAMPERS
ELECTRICAL AND
PNEUMATIC CONNECTIONS
BASE
PLATE
MAGNET BASE
MAGNET