80
ZTKS0174 / Z31966 / Rev.: 01
Technical Data MS 600‘89 Ascend
A.5.2 Technical Data Magnet
Operating modes of the Magnet System:
Driven Mode
In the driven mode the current lead is mounted and the electricity is flowing through the
power supply. The coils of the magnet can be energized and de-energized.
Persistent Mode
In the persistent mode the electricity is flowing exclusively within the magnet. The circuit
has no connection to the outside. The magnetic field cannot be switched off.
Technical Data Magnet
Value
Unit
Proton frequency
600
MHz
Central field
14.09
T
Coil inductance
49.3
H
Magnetic energy
1373.8
kJ
Magnetic center from bottom flange (design value)
435.2
mm
Maximum drift rate
0.01
ppm/h
6
Hz/h
Table A.9:
Specification of the Magnet
Summary of Contents for Ascend 600'89
Page 1: ...Version Innovation with Integrity NMR Spectroscopy 600 89 Ascend 01 User Manual ...
Page 4: ...4 ZTKS0174 Z31966 Rev 01 ...
Page 7: ...7 ZTKS0174 Z31966 Rev 01 ...
Page 8: ...8 ZTKS0174 Z31966 Rev 01 π ...
Page 10: ...10 ZTKS0174 Z31966 Rev 01 Contact ...
Page 34: ...34 ZTKS0174 Z31966 Rev 01 Transportation ...
Page 35: ...35 ZTKS0174 Z31966 Rev 01 4 Assembling 4 1 Safety Approved Persons Bruker Service only ...
Page 36: ...36 ZTKS0174 Z31966 Rev 01 Assembling ...
Page 57: ...57 ZTKS0174 Z31966 Rev 01 8 Disassembling 8 1 Safety Approved Persons Bruker Service only ...
Page 58: ...58 ZTKS0174 Z31966 Rev 01 Disassembling ...
Page 60: ...60 ZTKS0174 Z31966 Rev 01 Appendix ...
Page 62: ...62 ZTKS0174 Z31966 Rev 01 Appendix ...
Page 64: ...64 ZTKS0174 Z31966 Rev 01 Appendix ...
Page 66: ...66 ZTKS0174 Z31966 Rev 01 Appendix ...