53
ZTKS0175 / Z31991 / 03
Troubleshooting
6.3
Troubleshooting Work
6.3.1
After a Quench
Quench while in persistent mode:
1.
Wait until the quench valves are closed and no helium evaporates out of the quench
valves.
2.
Wait until there is no helium vapor visible in the room or the ceiling to make sure
there is sufficient oxygen in the room.
3.
If equipped with switch the Atmospheric Pressure Device OFF.
4.
Check the globes in the quench valves for their correct position.
5.
Only at magnet systems with 3 helium turrets: check the bursting disc at the helium
flow system. If destroyed close the socket with the supplied plug to prevent air from
entering the helium flow system or vessel.
6.
Remove probe and shim system to prevent icing of the shim system.
7.
Check the nitrogen turrets for icing.
8.
Start the refill with liquid helium as soon as possible after the quench (within one
hour after the quench; refer to the supplied Refilling Procedure).
9.
Contact Bruker Service immediately.
Figure 6.1:
Quench Picture
A quench is the very fast de-energizing of
the magnet by loss of its superconductivity.
The stored magnetic energy is converted
into heat, which promotes rapid evapora-
tion of large quantities of helium.
After an appearance of a quench contact
Bruker Service immediately.
WARNING
Cryogenic Agents
)
Quench
(see
)
Summary of Contents for 600'54 Ascend ULH
Page 1: ...Version Innovation with Integrity 600 54 Ascend ULH 03 User Manual NMR ...
Page 4: ...4 ZTKS0175 Z31991 03 ...
Page 7: ...7 ZTKS0175 Z31991 03 ...
Page 8: ...8 ZTKS0175 Z31991 03 π ...
Page 10: ...10 ZTKS0175 Z31991 03 Contact ...
Page 34: ...34 ZTKS0175 Z31991 03 Transportation ...
Page 35: ...35 ZTKS0175 Z31991 03 4 Assembling 4 1 Safety Approved Persons Bruker Service only ...
Page 36: ...36 ZTKS0175 Z31991 03 Assembling ...
Page 57: ...57 ZTKS0175 Z31991 03 8 Disassembling 8 1 Safety Approved Persons Bruker Service only ...
Page 58: ...58 ZTKS0175 Z31991 03 Disassembling ...
Page 60: ...60 ZTKS0175 Z31991 03 Appendix ...
Page 62: ...62 ZTKS0175 Z31991 03 Appendix ...
Page 64: ...64 ZTKS0175 Z31991 03 Appendix ...
Page 66: ...66 ZTKS0175 Z31991 03 Appendix ...
Page 68: ...68 ZTKS0175 Z31991 03 Appendix ...
Page 102: ...Bruker Corporation info bruker com www bruker com Order No ZTKS0175 Z31991 03 ...