Basic Setup Procedures
User Manual Version 002
BRUKER BIOSPIN
65 (327)
The figure above is an example of where the probe is either tuned to a completely
different frequency outside this window or the probe is not connected to the se-
lected preamp. Check
edasp
for correct routing, Check for correct matching box
frequency range. Increase WBSW to 50 or 100 and try to find the probe reso-
nance position.
When the probe is tuned well, as shown in
, start an acquisition by typ-
ing
zg
in the command line or clicking on the black triangle (upper left side) in the
acquisition display. Do a Fourier transformation and a phase correction by typing
ft
and phase correct.
Set offset
O1
to the value obtained for the center peak (see
) and
start
“
xau angle
”
. This will allow you to view the fourier transformed spectrum or
the FID after
ns
scans. The magic angle is adjusted best when the spikes on the
FID or the spinning sidebands in the
Spectrum
display have maximum size, like
shown in
. This is most easily seen with the carrier exactly on reso-
nance and the un-shuffled FID display mode. The spinning sidebands should
have maximum intensity, the rotational echoes on the FID should extend out to at
least 8msec in the FID-display.
Figure 4.9. FID and Spectrum of the
79
Br Signal of KBr used to Adjust the Magic
Angle
Calibrating 1H Pulses on Adamantane
4.3
Spin the KBr sample down and change to a spinner filled with adamantane. Spin
at 5-10 kHz. Generate a new data set from the KBr data set by typing
new
. Set
the instrument routing for
13
C observe and
1
H decoupling, as shown in the follow-
ing figure:
Summary of Contents for Solid State NMR
Page 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Page 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Page 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Page 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Page 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Page 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Page 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Page 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Page 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Page 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Page 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Page 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Page 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Page 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...