STMAS
User Manual Version 002
BRUKER BIOSPIN
247 (327)
ent case is in the order of MHz so even a small deviation causes a severe broad-
ening in the STMAS spectrum. This can easily be understood as the rotational
echoes decay much more rapidly when the magic angle is off. Experimentally it
has been found that the precision for setting the angle must be
≤
0.002°. The de-
pendence on the accuracy is so important that the experiment itself must be used
to find the most precise magic angle setting. It is obvious that this can only be
done on a well-known sample, as we will see later in the chapter. In order to
achieve the necessary precision for the adjustability of the angle a special goni-
ometer screw with an adapted gear transmission ratio is provided for the magic
angle setting knob as an upgrade for Bruker WB MAS probes. Once the best an-
gle setting is found it is advisable to leave the probe in the magnet. Sample
changes, however, on the WB probes doesn’t change the setting noticeably.
Pulse Sequences
19.2
and
show two of the basic sequences, which are 4-
pulse sequences with z-filter (
stmasdqfz.av
and
stmasdqfe.av
). Both sequences
start with a non-selective excitation pulse
p1
that creates SQ coherency on the in-
nermost ST which is allowed to evolve during the evolution period
D0
. Shortly be-
fore the end of the t
1
period there is a selective 180° pulse
P4
. This provides a
double quantum filter by which magnetization of the CT transition is eliminated
which will otherwise give a strong diagonal signal from a CT –> CT coherence
transfer pathway. The t
1
period is terminated with the second non-selective pulse
P2
.
Figure 19.2. Four-pulse sequence and coherence transfer pathway for the double
quantum filtered STMAS experiment with z-filter (stmasdqfz.av)
Pulses
p1
and
P2
are non-selective pulses. The corresponding power level
PL11
should be set to achieve around 100 kHz RF field amplitude.
P3
and
P4
are CT
selective pulse 90° and 180° pulses of about 20 and 40 µs, respectively, corre-
sponding to an RF field amplitude of a few kHz. Delays
D0
and
D4
are the incre-
mented delay for t
1
evolution and 20 µs for z-filter, respectively. Phase lists are as
follows, for phase sensitive detection in F1 the phase of the first pulse must be in-
cremented by 90° in States or States-TPPI mode:
Summary of Contents for Solid State NMR
Page 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Page 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Page 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Page 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Page 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Page 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Page 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Page 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Page 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Page 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Page 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Page 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Page 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Page 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...