42 (47)
BRUKER BIOSPIN
Operating & Service Manual Version 001
Service Information and Maintenance
Figure 8.1. Fan Supplies and Status Connections
4. Unscrew only the 4 screws from the top of the Push and Pull fan assemblies
on the both sides of the RF module.
Figure 8.2. Push Fan Assembly
5. Remove the Push and Pull fan assemblies.
6. Place correctly the new fan assembly sets in the bottom holes of the RF
module and screw it on the top.
7. Connect the 2 red/white wires on the left side of the RF module (J3 and J4).
8. Connect line cord and turn on the amplifier. Note that the fans are turning and
no OVERHEAT status led appears on front panel.
9. Turn off the amplifier and disconnect the line cord.
10. Put the coverage plate on the amplifier and screw it.
11. Put the amplifier in the CLINSCAN console, connect all cables on the front
panel and the line cord on the rear panel.
Pull fan
connector J3
Push fan
connector J4
Screws for Push an Pull
fan assemblies