PULSE Multi-analyzer System Type 3560-B/C/D/E –
Installation and IDAe Hardware
138
A.2
Connection of Synchronization Cables
IDA BNT
Using the IDA BNT synchronization you need two cables such as AO-0158, a balanced
cable fitted with BNT connectors.
Connect the Sync. Out and the Clock Out on the master to the Sync. In and Clock In. on the
second frame as shown in Fig.A.2.
The switch marked
Int.-Ext
. must be in position
Int.
on the master and
Ext.
on the second
frame.
Fig.A.2
Connecting multiple
frames to form a front-end
using IDA/BNT
When connecting a third frame, connect the Clock and the Sync. Out signals from the
second frame to the Clock and Sync. Input on the third frame. The switch on the third frame
must be in position
Ext.
The cables used for the Clock and the Sync. Signals must be triaxial cables AO-0158 or
equivalent. The cables should have the same length; the maximum length is 100 m.
Next, connect the LAN cables using a switch as described in section 2.3.1.
020245
Summary of Contents for PULSE 3560-B
Page 2: ......
Page 8: ......
Page 11: ...Part I Installation...
Page 12: ......
Page 44: ...PULSE Multi analyzer System Type 3560 B C D E Installation and IDAe Hardware 36...
Page 53: ...Part II Hardware...
Page 54: ......
Page 110: ...PULSE Multi analyzer System Type 3560 B C D E Installation and IDAe Hardware 102...
Page 140: ...PULSE Multi analyzer System Type 3560 B C D E Installation and IDAe Hardware 132...
Page 142: ...PULSE Multi analyzer System Type 3560 B C D E Installation and IDAe Hardware 134...
Page 143: ...Part III Appendices...
Page 144: ......
Page 148: ...PULSE Multi analyzer System Type 3560 B C D E Installation and IDAe Hardware 140...
Page 161: ......